Connect a suitable tubing to a Column valve port, for example port 2A if
column position 2 was chosen in the method to be run.
4
Open the System Control module and select Manual:Execute Manual
Instructions.
5
Result: The Manual instructions dialog opens.
In the Manual instructions dialog:
6
•
Select Flow path:Column position.
•
Choose the Position used for connection of the column, in this example
Position 2.
•
Choose the Flow Direction Down flow.
•
Click
.
Result: The Column valve switches to position 2.
Getting Started with ÄKTAavant and UNICORN 6 28-9440-69 AB
85
7 Prepare the system for a run
7.4 Connect a column