TIP:
Empty the capillary loop with a buffer volume that exceeds the volume
of the loop. This will ensure that the loop is completely emptied.
How to fill a capillary loop
Follow the instruction below to fill the capillary loop with sample.
Action
Step
Connect a suitable capillary loop to Injection valve ports LoopF (fill) and
LoopE (empty).
1
Fill a syringe with sample.
2
Connect the syringe to the Injection valve port Syr.
3
Getting Started with ÄKTAavant and UNICORN 6 28-9440-69 AB
99
8 Run a method
8.2 Applying the sample