2 Preinstallation instructions
32551-IMG rev 6
GE Healthcare
15
2.11 U
NPACKING
THE
C-
ARM
– cut the packing straps
– unscrew Phillips screws
– open the fastened cover
– lift out the security box with the C-arm
– remove the supporting bracket by loosening the bolt inside the
crate
– remove the nuts securing the metal packing bracket to the crate
– lift out the C-arm
– remove the packing bracket from the C-arm.
CAUTION!
The bolts holding the bracket are not meant to be used as attachment
bolts (they are too long). Attachment bolts are supplied separately in a
plastic bag.
Fig 2.3.
C-arm packing
2.12 U
NPACKING
THE
COUNTERWEIGHT
– cut the packing straps
– unscrew Phillips screws
– open the fastened cover
– remove the spacing boards and lay them next to the crate
– turn the crate over onto the spacing boards
– lift the empty crate off the counterweight