4 Alpha RT softstart procedure
32551-IMG rev 6
GE Healthcare
37
4 Alpha RT softstart procedure
4.1
P
REREQUISITES
Turn the power off from the main power switch.
CAUTION!
The purpose of this procedure is to ensure that before applying full power
to the generator, all vital signals and adjustments are correct. Use large
focus (“MAGN” lamp is off).
WARNING!
When installing or removing covers, make sure that no articles such
screws etc. can fall through to the open spaces between the column and
the carriage. Serious problems for the use of the device may occur.
4.2
R
EMOVAL
OF
THE
SIDE
COVERS
1
Remove the side and back covers to expose the circuit boards.
CAUTION!
Do not pull out the side cover, just straight down.
CAUTION!
Side cover tabs are sharp.
Open spaces
between the column
and the carriage