5 Setups, alignments and adjustments
48
GE Healthcare
32551-IMG rev 6
The kV is selected on the basis that the target time for the compressed
breast thickness will be reached. The higher target time will result in lower
selected kV and vice versa.
To enter the service mode:
To enter the service mode press
kV-
and
density-
buttons simultaneously
until the Timer display shows
Ser
.
To enter the special program menu:
To enter the special program menu press
and
density-
buttons
simultaneously. The program number is displayed in the kV display and
the program name in the mAs display (Press DEN- first because otherwise
the system may toggle the focal spot with DEN+ key). The available
programs can be scrolled with the
kV+
and
kV-
buttons. To enter the
desired program press the
Autoexp
program button. To leave the
program press
Autoexp
buttons once more. To leave the menu press
and
density-
simultaneously.
Return to the menu from the special programs by pressing
AUTOEXP
button.
NOTE!
See block diagram of software in appendix or service manual.
5.1.2
Setup Info Printout and printer/PC connection #62 InF
Connect a matrix printer or a PC/Laptop to Alpha RT com port. Select
service program #52SUP with the
kV+
or
kV-
buttons. Enter the program
pressing the
Autoexp
button. Press
Autoexp
to exit program #52SUP.
Select service program #62Inf with the
kV+
or
kV-
buttons. Press
Autoexp
to enter the program. The info is printed out immediately. See an example
of the printout on next page.
The serial connection is an ordinary asynchronous receiver/transmitter
(RS-232). The serial port is initialized in Alpha RT as follows:
– 1200 baud if printer or Alpha ID II is connected
– 9600 baud if Dataview (or Alpha ID) is connected (in
S
ET
UP
parameter i =2)
– 8,N,1 (8 databits, No parity [always 0], 1 stop bit)
The effect of the ‘no parity bit’ and the stop bit is a total of two stop bits, so
receiving end must be configured correspondingly.
The data bytes are:
The data can be received with a communications program, for example
Windows (Hyper)Terminal, to PC and printed afterwards on paper, or a
printer can be directly connected to the port.
start 8-bit data 1 stop
Summary of Contents for Alpha RT 32551
Page 1: ...GE Healthcare Alpha RT Installation Instructions 32551 IMG rev 6 0459 ...
Page 2: ......
Page 4: ......
Page 42: ...3 Installation 36 GE Healthcare 32551 IMG rev 6 ...
Page 51: ...4 Alpha RT softstart procedure 32551 IMG rev 6 GE Healthcare 45 Fig 4 10 Inverter board ...
Page 52: ...4 Alpha RT softstart procedure 46 GE Healthcare 32551 IMG rev 6 ...
Page 71: ...5 Setups alignments and adjustments 32551 IMG rev 6 GE Healthcare 65 Fig 5 2 Magnetic brake ...
Page 94: ...5 Setups alignments and adjustments 88 GE Healthcare 32551 IMG rev 6 ...
Page 96: ...6 Alpha RT installation step by step 90 GE Healthcare 32551 IMG rev 6 ...
Page 97: ......
Page 98: ......