5 Setups, alignments and adjustments
32551-IMG rev 6
GE Healthcare
55
#53PAr/b The density display shows
b
which means the
Anode Brake
Mode
. Select 0 or 1 using the Den+ and Den- buttons.
0 = Anode Brake off
1 = Anode Brake on
The factory default is to have the anode brake off after an
exposure. This reduces the heat loading of the tube housing
and applies less stress to the bearings.
Press
kV+
#53PAr/t The density display shows
t
. In this mode the software for
Alpha RT is selected.
0 = Not for Alpha RT
1 = Not for Alpha RT
2 = Only AA and manual mode (Alpha RT)
3 = Only A and manual mode (Alpha RT)
4 = AA ->> A ->> manual mode (factory default
Alpha RT)
5 = A ->> AA ->> manual mode (Alpha RT)
Press
kV+
#53PAr/I Collimator type (Interlock)
1 for MGB-500 Collimator (with interlock)
0 for MGB-400 Collimator (no interlock)
Press
kV+
#53PAr/d C-arm control board selection
0 = C-arm control board (32310)
1 = CD-board (32010)
Press
Autoexp
to leave the
PAr
program.
Press
kV+
Summary of Contents for Alpha RT 32551
Page 1: ...GE Healthcare Alpha RT Installation Instructions 32551 IMG rev 6 0459 ...
Page 2: ......
Page 4: ......
Page 42: ...3 Installation 36 GE Healthcare 32551 IMG rev 6 ...
Page 51: ...4 Alpha RT softstart procedure 32551 IMG rev 6 GE Healthcare 45 Fig 4 10 Inverter board ...
Page 52: ...4 Alpha RT softstart procedure 46 GE Healthcare 32551 IMG rev 6 ...
Page 71: ...5 Setups alignments and adjustments 32551 IMG rev 6 GE Healthcare 65 Fig 5 2 Magnetic brake ...
Page 94: ...5 Setups alignments and adjustments 88 GE Healthcare 32551 IMG rev 6 ...
Page 96: ...6 Alpha RT installation step by step 90 GE Healthcare 32551 IMG rev 6 ...
Page 97: ......
Page 98: ......