2 Preinstallation instructions
10
GE Healthcare
32551-IMG rev 6
2.2
S
TRENGTH
/
DURABILITY
REQUIREMENTS
FOR
THE
WALL
AND
FLOOR
If the device is mounted on a wall, the pulling force exerted on each bolt is
approximately 600 N (135 lbs). The load on the floor is approximately 7350
kg/sq.m (1500 lb/sq.ft).
If the device is mounted on a base plate, the load on the floor is
approximately 600 kg/sq.m (122 lb/sq.ft).
2.3
E
LECTRICAL
SUPPLY
REQUIREMENTS
NOTE!
Operating Range 198 to 242VAC. Line voltages measuring less than 208
VAC may require a line boost transformer.
Line Voltage
220 VAC (nominal)
(tap selection 220, 240 or 260 VAC)
Line Regulation
± 2% (full load)
Phase
single phase (Line, Neutral,
Protective Ground or L1, L2,
Protective Ground)
Frequency
50/60Hz
Current
stand-by 1A
maximum 20 A
instantaneous peak 30 A
Branch Circuit: dedicated Line,
Neutral, Protective Ground 2.5mm2
(L1 and L2 10 AWG, stranded wire,
continuous run)
circuit breaker 20 A, 2-pole
(outlet per local code requirements
or NEMA 6-20)
VDE Wall plug is included
Grounding
dedicated, wired (not conduit)
0-potential (earth) ground
Standard outlet
(220/120V/15A) required for any
optional accessories
Summary of Contents for Alpha RT 32551
Page 1: ...GE Healthcare Alpha RT Installation Instructions 32551 IMG rev 6 0459 ...
Page 2: ......
Page 4: ......
Page 42: ...3 Installation 36 GE Healthcare 32551 IMG rev 6 ...
Page 51: ...4 Alpha RT softstart procedure 32551 IMG rev 6 GE Healthcare 45 Fig 4 10 Inverter board ...
Page 52: ...4 Alpha RT softstart procedure 46 GE Healthcare 32551 IMG rev 6 ...
Page 71: ...5 Setups alignments and adjustments 32551 IMG rev 6 GE Healthcare 65 Fig 5 2 Magnetic brake ...
Page 94: ...5 Setups alignments and adjustments 88 GE Healthcare 32551 IMG rev 6 ...
Page 96: ...6 Alpha RT installation step by step 90 GE Healthcare 32551 IMG rev 6 ...
Page 97: ......
Page 98: ......