5 Setups, alignments and adjustments
32551-IMG rev 6
GE Healthcare
73
5.2.9
kV, mAs and exposure time test
The measurement must be performed with a calibrated RTI, PMX-I/M or
equivalent.
1
Go into
service mode
when performing the test. Select
manual
mode
.
2
Attach
cassette tunnel
when measuring the
LF
. The Syncro Bucky
feature interrupts the x-rays for 50ms at grid turnaround points. The
interruption interferes the measurement.
When measuring the
SF, no cassette holder
should be used. Set the
parameters as shown in the table below and make the exposures.
3
Record the kV, mAs and time reading after each exposure. The time
reading is to be compared with the Alpha RT control panel time
display.
The maximum tolerance for kV, mAs and exposure time are:
For kV
± ( 1.5 kV + instrument inaccuracy) =
± ( 1.5 kV + 1.5 % of r 600V)
For mAs
± 4 mAs for mAs values less than or equal to 40 mAs
± 10 % for mAs values greater than 40 mAs
For exposure time
± ( 0.1 instrument inaccuracy ) =
± ( 0.1 0.25 msec + 0.5 % of reading
kV adjustment
kV can be adjusted from R 63 located on the Inverter Board. See
kV
mAs
kV
mAs
kV
mAs
LF
23
300
28
125
35
63
SF
28
63
35
20
Table 5.6 Parameters for small and large focuses
Summary of Contents for Alpha RT 32551
Page 1: ...GE Healthcare Alpha RT Installation Instructions 32551 IMG rev 6 0459 ...
Page 2: ......
Page 4: ......
Page 42: ...3 Installation 36 GE Healthcare 32551 IMG rev 6 ...
Page 51: ...4 Alpha RT softstart procedure 32551 IMG rev 6 GE Healthcare 45 Fig 4 10 Inverter board ...
Page 52: ...4 Alpha RT softstart procedure 46 GE Healthcare 32551 IMG rev 6 ...
Page 71: ...5 Setups alignments and adjustments 32551 IMG rev 6 GE Healthcare 65 Fig 5 2 Magnetic brake ...
Page 94: ...5 Setups alignments and adjustments 88 GE Healthcare 32551 IMG rev 6 ...
Page 96: ...6 Alpha RT installation step by step 90 GE Healthcare 32551 IMG rev 6 ...
Page 97: ......
Page 98: ......