6 Alpha RT installation step by step
32551-IMG rev 6
GE Healthcare
89
6 Alpha RT installation step by step
1
Select the
default kV values
that the site will use with the bucky,
cassette tunnel and magnification table. Mark the values in the
tables 5.8, Alpha RT settings for Molybdenum filter and 5.9, Alpha RT
2
Select the
nearest film & screen combination
Alpha RT settings for Molybdenum filter that the site will use with the
bucky, cassette tunnel and magnification table. Mark the values in
the tables 5.8, Alpha RT settings for Molybdenum filter and 5.9, Alpha
RT settings for Rhodium filter.
3
Select molybdenum filter and install the bucky.
Adjust the AEC
software gain
. See section
AEC Adjustment and Thickness
compensation adjustment
.
See also AEC adjustment quick fix guide.
4
Program the reference value for the IQC. See section
Pr#64 AEC,
parameters d, h
and Operation Instructions section
INTEGRATED
QUALITY CONTROL TEST
.
5
Go into (A) mode. Install the magnification kit. Make an exposure with
40mm acrylic using 30 kV or a clinical kV at that thickness. Process
the film and measure the optical density.
– if the optical density is too high, select density- until the wanted
optical density is reached.
– if the optical density is too low, select until the wanted
optical density is reached.
Mark the new density value in the tables 5.8, Alpha RT settings for
Molybdenum filter and 5.9, Alpha RT settings for Rhodium filter.
Repeat this test with the cassette tunnel.
6
Select rhodium filter and repeat the previous test for the
bucky,
cassette holder and magnification tunnel
.
7
Go into program (Pr) mode (press Den-&Den+) and program the
default values marked in the tables 5.8, Alpha RT settings for
Molybdenum filter and 5.9, Alpha RT settings for Rhodium filter.
8
Select program #62 INF and press Autoexp to print the programmed
values.
6.1
S
OFTWARE
BLOCK
DIAGRAM
See appendix.
Summary of Contents for Alpha RT 32551
Page 1: ...GE Healthcare Alpha RT Installation Instructions 32551 IMG rev 6 0459 ...
Page 2: ......
Page 4: ......
Page 42: ...3 Installation 36 GE Healthcare 32551 IMG rev 6 ...
Page 51: ...4 Alpha RT softstart procedure 32551 IMG rev 6 GE Healthcare 45 Fig 4 10 Inverter board ...
Page 52: ...4 Alpha RT softstart procedure 46 GE Healthcare 32551 IMG rev 6 ...
Page 71: ...5 Setups alignments and adjustments 32551 IMG rev 6 GE Healthcare 65 Fig 5 2 Magnetic brake ...
Page 94: ...5 Setups alignments and adjustments 88 GE Healthcare 32551 IMG rev 6 ...
Page 96: ...6 Alpha RT installation step by step 90 GE Healthcare 32551 IMG rev 6 ...
Page 97: ......
Page 98: ......