Condition
Parameter
Front of instrument (the operation panel side): 40 cm
Free space required around
Amersham Imager 600
Right: 10 cm
Left: 10 cm
Rear: 10 cm
Top: 10 cm
A stable laboratory bench with a load capacity of 980 N (100 kg) or
higher
Placement
The bench must be level within 2º (two) degrees.
1
Take into consideration the workflow and ancillary facilities when
planning the installation.
2
Any required construction and electricity/air conditioning work
must be completed prior to installation.
3
It is not desirable to have a heat source near the Amersham
Imager 600 air intakes, even if the other environmental require-
ments have been met.
4
Do not install the equipment near a window. Avoid direct sunlight.
Ensure blinds are attached to nearby windows.
5
Do not place objects near the power outlet to ensure easy access
to the power cord for disconnection in case of emergency.
Other conditions
Temperature: 18°C to 28°C (with temperature fluctuation below 10ºC
per hour or lower)
Operating temperature/humidity
conditions
Humidity: 20% to 70% RH (no dew condensation)
Note:
When the above conditions cannot be satisfied, take appropriate
actions.
Temperature: -25°C to 70°C
Transportation/storage
conditions
Humidity: 5% to 95% RH (no dew condensation)
Note:
When the above conditions cannot be satisfied, take appropriate
actions.