Condition
Parameter
Do not install the instrument:
Installation location conditions
1
where the temperature can vary widely (with temperature fluc-
tuation above 10ºC per hour)
2
near a heat source such as a radiator
3
where it may get wet or flooded.
4
where it may be exposed to corrosive gas
5
in a dusty environment
6
in a place constantly or excessively exposed to vibration or im-
pacts
7
in a place exposed to direct sunlight
Indoor use
Operation site
2000 m or lower
Maximum operating altitude
70 dB(A) or lower
Noise
Fast Lmax ≤60dB(A), and average Leq ≤ 54dB(A) (1 m from the instru-
ment)
IP21
Protection provided by enclosure
Transient overvoltage category II
Overvoltage category
Pollution degree 2
Rated pollution applied
Amersham Imager 600 Operating Instructions 29-0645-17 AD
223
9 Specifications