5.3
Select tray
The following table describes the suitable tray and accessory for each capturing method.
Use the correct tray to the selected exposure for optimal image quality.
Note:
Available in
configura-
tion
Tray and accessory
Sample type
Image capturing method
600
Black Tray
•
Membrane
Chemiluminescence
600 UV
600 QC
600 RGB
600
Black Tray +
•
Membrane
Chemiluminescence
with colorimetric marker
600 UV
White Insert
600 QC
600 RGB
600
Black Tray +
•
Membrane
•
Gel
Colorimetric
Epi-illumination
600 UV
600 QC
White Insert or,
White Trans Tray
600 RGB
600 QC
White Trans Tray +
•
Gel
Colorimetric
Trans-illumination
600 RGB
Diffuser Board
600 RGB
Black Tray
•
Gel
•
Membrane
Fluorescence RGB
Epi-illumination
600 UC
UV Trans Tray
(transparent) +
•
Gel
Fluorescence
Trans-UV
600 QC
Gel sheet
600 RGB
NOTICE
With regards to third party patents related to sample preparation
etcetera, we recommend that you consult a lawyer or patent attor-
ney on obtaining appropriate licenses.