14
OPM_ITS_USM_M06_M20_0US_V011 (09/2007)
•
When replacing batteries, replace with
12Volt/34Watt cell and exact quantity as listed below based on UPS VA rating:
•
600VA – 2 batteries
•
1000VA – 2 batteries
•
1500VA – 3 batteries
•
2000VA – 4 batteries
•
Please call your local GE representative for pricing and availability.
9.5 Hot Swappable Battery Replacement
Tower Procedure -
Step
1.
Remove front cover panel; pull out from top where finger grooves are located on each side. (See diagrams below)
600 & 1000VA
1500 & 2000VA
Step 2.
Push down slightly to allow front cover to be moved to the side. (See diagrams below)