4.1
Site requirements
Environmental conditions
The following table describes the environmental requirements for Amersham Imager
600.
CAUTION
Do not use the instrument in a room with a temperature above
+28°C. Higher temperatures does not allow the CCD to cool down
properly (to -25°C).
Condition
Parameter
Front of instrument (the operation panel side): 40 cm
Free space required around
Amersham Imager 600
Right: 10 cm
Left: 10 cm
Rear: 10 cm
Top: 10 cm
A stable laboratory bench with a load capacity of 980 N (100 kg) or
higher
Placement
The bench must be level within 2º (two) degrees.
1
Take into consideration the workflow and ancillary facilities when
planning the installation.
2
Any required construction and electricity/air conditioning work
must be completed prior to installation.
3
It is not desirable to have a heat source near the Amersham
Imager 600 air intakes, even if the other environmental require-
ments have been met.
4
Do not install the equipment near a window. Avoid direct sunlight.
Ensure blinds are attached to nearby windows.
5
Do not place objects near the power outlet to ensure easy access
to the power cord for disconnection in case of emergency.
Other conditions
Amersham Imager 600 Operating Instructions 29-0645-17 AD
63
4 Installation
4.1 Site requirements