5.3
Select tray
The following table describes the suitable tray and accessory for each capturing method.
Use the correct tray to the selected exposure for optimal image quality.
Note:
Available in
configura-
tion
Tray and accessory
Sample type
Image capturing method
680
Black Tray
•
Membrane
Chemiluminescence
680 UV
680 QC
680 RGB
680
Black Tray +
•
Membrane
Chemiluminescence
with colorimetric marker
680 UV
White Insert
680 QC
680 RGB
680
Black Tray +
•
Membrane
•
Gel
Colorimetric
Epi illumination
680 UV
680 QC
White Insert or,
White Trans Tray
680 RGB
680 QC
White Trans Tray +
•
Gel
Colorimetric
Trans-illumination
680 RGB
Diffuser Board
680 RGB
Black Tray
•
Gel
•
Membrane
Fluorescence RGB
Epi-illumination
680 UV
UV Trans Tray
(transparent) +
•
Gel
Fluorescence
Trans-UV
680 QC
Gel sheet
680 RGB
NOTICE
With regards to third party patents related to sample preparation
etcetera, it is recommended that you consult a lawyer or patent
attorney for information on obtaining appropriate licenses.
Amersham Imager 680 Operating Instructions 29271308 AB
105
5 Operation
5.3 Select tray