8.2
Error messages
While starting the instrument
Corrective action
Meaning
Error message
Check that the scanner is turned
on and connected to the comput-
er.
The scanner is not detected.
Cannot detect Amersham Ty-
phoon scanner. Please check
connection and power.
While preparing for a scan
Corrective action
Meaning
Error message
Close the scanner door.
The scanner door is open
Close the scanner door.
Close the filter door.
The filter door is open.
Close the filter door.
Select an image folder by click-
ing the Browse... button and
navigating to a suitable folder.
No image folder is selected.
Please select image folder.
Type a name in the File Name
field.
No name has been assigned to
the image file.
Please input a file name.
Type a PMT voltage value be-
tween 250 and 1000 in the PMT
field.
The PMT voltage setting is out-
side the permitted range.
Please set PMT voltage in the
range 250 to 1000.
Insert the stage properly.
The stage is not properly insert-
ed.
The stage is not properly insert-
ed. Please insert the stage in its
correct position.
During a scan
Corrective action
Meaning
Error message
Decrease the PMT voltage in the
Scan settings pane.
The scanned image was overex-
posed due to a high PMT voltage
setting.
The scan was stopped because
of overexposure. Set a lower
PMT voltage and restart the
scan.
Amersham Typhoon Operating Instructions 29193226 AC
129
8 Troubleshooting
8.2 Error messages