Model T9 Ultrasonic Flow Transducer Installation Guide
23
Chapter 3. Inserting T9 Transducers into the Pipe
3.4 Inserting Transducers with the Flanged Holder (cont.)
3.
Tighten the compression fitting to hold the transducer in place.
4.
Place a spiral-wound gasket over the end of the transducer.
Note:
You must use a gasket in order to provide the proper acoustic isolation.
5.
Inspect the pipe nozzle. Make sure it is free from dirt/rust. Use the steel wool to clean the pipe nozzle if necessary.
6.
Insert the transducer into the nozzle.
Gasket
Gasket