Programming Site Data
2-13
September 2004
Setting Up the External
Totalizer Reset Switch
If you have installed an external totalizer reset switch as described in
Chapter 1,
Installation
, the
GATE OPTION
enables you to configure
the switch.
1.
At the
GATE OPTION
prompt, press [
X
] to select one of the
following and press
[ENTER]
.
•
RESET
- the meter resets totals to zero for both channels and
then immediately begins to count again.
•
HOLD
- the meter holds totals at the current point for both
channels. The meter will not count as long as the switch
remains closed. This option is typically used for calibration.
2.
Do one of the following:
•
If MASS FLOW is ON and both channels are active, proceed to
Selecting Mass Flow Units
below.
•
If MASS FLOW is OFF or only one channel is active, the meter
returns to the
Global PROGRAM
menu. Go to
Settings Up the
Outputs
on the next page.
Selecting Mass Flow Units
1.
Press [
W
]
and [
X
] to select the desired mass flow units for the
flow rate display and press
[ENTER]
. The available units for this
prompt are determined by the selection made at
SYSTEM UNITS
on
the previous page.
2.
Press [
W
]
and [
X
] to select the desired time units for the mass
flow rate display and press
[ENTER]
.
3.
Press [
W
]
and [
X
] to select the desired number of digits to the
right of the decimal point in the mass flow rate display and press
[ENTER]
.
4.
Press [
W
]
and [
X
] to select the desired units for the totalized mass
flow rate display and press
[ENTER]
. The available units for this
prompt are determined by the selection made at
SYSTEM UNITS
on
the previous page.
5.
Press [
W
]
and [
X
] to select the desired number of digits to the
right of the decimal point in the totalized mass flow rate display
and press
[ENTER]
.