which will move the display to the PROBE fi eld ready
to select the code number for the probe being set.
When the code number has been incremented to the
correct one, a press on the
MENU
button will take the
instrument back to the normal measurement display
which has the code indicated on the bottom line and
will show ERROR (see Figure below).
– To re-commence the set probe routine, once the
correct probe code has been entered, press
SET
.
5.4
Entering reference block
values
Before the AutoSigma 3000 DL can be set, the ref-
erence block values need to be entered. It is recom-
mended that the STD blocks supplied with the unit are
used as these will provide the most effective and ac-
curate setting. As an alternative user reference blocks
may be used, these will limit the instruments range
and can reduce accuracy.
Ensure that the instrument is set to the correct units
for the block values to be entered. If not change the
UNITS parameter value fi eld before moving to the STD
parameters and entering the block values. From the
normal measurement mode proceed as follows, e.g.
setting STD 1 (60.1) and STD 2 (8.91).
Changing working parameters
Operation
5 - 4
Issue 04, 06/2007
AutoSigma 3000
Summary of Contents for AutoSigma 3000 DL
Page 1: ...AutoSigma 3000 Technical Reference and Operation Manual Ident No 071 002 022 47DH01 105 952...
Page 6: ...0 6 Issue 04 06 2007 AutoSigma 3000...
Page 7: ...Introduction 1 AutoSigma 3000 Issue 04 06 2007 1 1...
Page 12: ...1 6 Issue 04 06 2007 AutoSigma 3000...
Page 13: ...Standard package and accessories 2 AutoSigma 3000 Issue 04 06 2007 2 1...
Page 16: ...2 4 Issue 04 06 2007 AutoSigma 3000...
Page 17: ...Preparation for operation 3 AutoSigma 3000 Issue 04 06 2007 3 1...
Page 27: ...Basic functions 4 AutoSigma 3000 Issue 04 06 2007 4 1...
Page 31: ...Operation 5 AutoSigma 3000 Issue 04 06 2007 5 1...
Page 47: ...External data handling devices 6 AutoSigma 3000 Issue 04 06 2007 6 1...
Page 52: ...6 6 Issue 04 06 2007 AutoSigma 3000...
Page 53: ...Specifications 7 AutoSigma 3000 Issue 04 06 2007 7 1...
Page 57: ...Maintenance and care 8 AutoSigma 3000 Issue 04 06 2007 8 1...
Page 59: ...Appendix 9 AutoSigma 3000 Issue 04 06 2007 9 1...
Page 62: ...9 4 Issue 04 06 2007 AutoSigma 3000...
Page 63: ...Index 10 AutoSigma 3000 Issue 04 06 2007 10 1...