Sending serial data to an external device
6.2 Sending serial data to an
external device
Connect the external equipment to the instrument via
the correct interface cable and switch it on.
The AutoSigma 3000 DL will recognise the connection
and change the use of the
SEND
button from internal
data logging to external serial data transfer.
Individual fi les or the entire memory can be transmit-
ted, but only fi les containing stored data will be trans-
ferred i.e. fi le active or fi le full.
A long press on the
SEND
button is used to initiate
data out transfer with the File menu set to a single fi le
or ALL FILES.
H
Note:
For correct handling of transmitted data PCs and print-
ers should be set to Xon / Xoff protocol.
Sending a single fi le
With the instrument in the Measurement mode:
Step
Key
Display
1
Open the fi le menu.
2
If fi le to be transferred is
not current (i.e. the dis-
played one), select the
fi le required (e.g.
2
).
3
Transfer the fi le with
a long press; after the
press the display will in-
dicate the transfer.
When transmission is complete the display will revert
to the fi le selection.
H
Note:
To terminate a data transfer when in progress press
SEND
. Display will respond for a short period with
External data handling devices
AutoSigma 3000
Issue 04, 06/2007
6 - 3
Summary of Contents for AutoSigma 3000 DL
Page 1: ...AutoSigma 3000 Technical Reference and Operation Manual Ident No 071 002 022 47DH01 105 952...
Page 6: ...0 6 Issue 04 06 2007 AutoSigma 3000...
Page 7: ...Introduction 1 AutoSigma 3000 Issue 04 06 2007 1 1...
Page 12: ...1 6 Issue 04 06 2007 AutoSigma 3000...
Page 13: ...Standard package and accessories 2 AutoSigma 3000 Issue 04 06 2007 2 1...
Page 16: ...2 4 Issue 04 06 2007 AutoSigma 3000...
Page 17: ...Preparation for operation 3 AutoSigma 3000 Issue 04 06 2007 3 1...
Page 27: ...Basic functions 4 AutoSigma 3000 Issue 04 06 2007 4 1...
Page 31: ...Operation 5 AutoSigma 3000 Issue 04 06 2007 5 1...
Page 47: ...External data handling devices 6 AutoSigma 3000 Issue 04 06 2007 6 1...
Page 52: ...6 6 Issue 04 06 2007 AutoSigma 3000...
Page 53: ...Specifications 7 AutoSigma 3000 Issue 04 06 2007 7 1...
Page 57: ...Maintenance and care 8 AutoSigma 3000 Issue 04 06 2007 8 1...
Page 59: ...Appendix 9 AutoSigma 3000 Issue 04 06 2007 9 1...
Page 62: ...9 4 Issue 04 06 2007 AutoSigma 3000...
Page 63: ...Index 10 AutoSigma 3000 Issue 04 06 2007 10 1...