Parameters
5-21
Document no. 2044677-001
4.4.1 NIBP Calibration
How to calibrate
NOTE: Perform NIBP
Calibration Check
first to evaluate if calibration is needed or not.
NOTE: Both channels B1 and B2 must be calibrated simultaneously.
NOTE: The module must be in the frame during the whole procedure.
NOTE: Calibration selection is available only when protection is OFF.
1.
Press the NIBP module buttons
Auto ON/OFF
and
Start Cancel
simultaneously for 3
seconds to enable the
Protection
selection - the color of the
Protection
selection
changes from grey to white.
NOTE: Make sure that you are using the module keys and not the side panel keys of the
monitor.
NOTE: When the buttons have been pressed, the message ‘Calibration switch on!’ is shown in
the NIBP digit field.
2.
Change the
protection
setting from ON to OFF to enable the
Calibrate
selection - the
color of the
Calibrate
selection changes from grey to white.
NOTE: When calibration has been enabled, the message ‘Calibration not protected’ is shown in
the NIBP digit field.
3.
Zeroing:
−
Disconnect the NIBP hose from the module connector.
−
Select
Calibrate
and push the
ComWheel.
Active
Leak
Test
:
Wrap an adult cuff around a pipe
and connect the cuff to the module. Select the
active leak test (ON). The module automatically
pumps a pressure of 280 mmHg into the cuff. Wait
for several seconds until the pressure stabilizes.
Then check that the pressure reading does not drop
more than 5 mmHg per minute. If it does, leaking
point(s) should be detected and fixed. Cancel the
test by selecting the Active leak test OFF.
Calibration Check
:
After the calibration check is
selected , the module zeroes the pressure
transducers at the beginning of the calibration
check. Do not pump pressure until the text
‘Calibrating’ appears in the NIBP digit field or the
zeroing will fail. After the zeroing is done, manually
pump pressure into the module and make sure that
the same pressure values are shown both on the
display and on the manometer. Pressure of both
pressure channels B1 and B2 are shown. The
pressure values are automatically
zero-compensated, so the readings of B1 and B2 should be the same as the manometer readings.
Protection
:
Software calibration protection (ON/OFF). Select
OFF
when calibrating.
Protection
selection
becomes available in the menu after pressing the NIBP module buttons
Auto ON/OFF
and
Start/Cancel
simultaneously for 3 seconds.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......