Airway gas (CO2)
17-5
CO
2
setup menu
Scale
Allows you to select the following scales:
FiO2 Level
Selects the FiO
2
level. The FiO
2
level is used in gas compensations of the CO
2
measurement to
increase the measurement accuracy. Make sure the level is set if the FiO
2
level is higher than
40%.
Unit
Selects the CO
2
display unit:
%
,
kPa
or
mmHg
.
Resp Rate Source
You can select the rate to be calculated from the ECG leads (impedance measurement) or the
CO
2
measurement. If you select AUTO, the rate is automatically calculated from the measured
CO
2
. If impedance measurement is present and the source is AUTO, the respiration rate
calculation switches back to impedance respiration if you press the
Silence Alarms
key
during an Apnea alarm.
CO2 Alarm
Opens the CO
2
Alarms Adjustment
menu to change the CO
2
alarm limits.
Resp Rate Alarm
Opens the
Resp Rate Alarms Adjustment
menu to change the respiration rate alarm limits.
Previous Menu
Returns to the previous menu.
Selecting alarm sources
To select alarm sources, go to the
CO2 Setup
menu’s submenu
CO2 Alarm
. The selections are
FI
or
ET
as the high and low alarm limit.
scale options for %
scale options for kPa
scale options for mmHg
0-6%
0-10%
0-15%
0-20%
0 - 6 kPa
0 - 10 kPa
0 - 15 kPa
0 - 20 kPa
0 - 50 mmHg
0 - 80 mmHg
0 - 100 mmHg
0 - 160 mmHg
Airway
Gas
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......