background image

7  |  GE Oil & Gas

© 2014 General Electric Company. All rights reserved.

3�   Remove the positioner from the bracket� The tubing and 

fittings on each side of the positioner must be taken off and 

installed in the opposite corner from their original position 

(see piping schematic provided by GE), then reattach to 

the positioner� This will allow the positioner to maintain the 

actuator failure mode when the positioner is turned upside 

down�

4�   Move the bracket assembly, tube assembly, and cover plate 

to the positions shown on the desired drawing�

5�   Install the positioner in the opposite direction of its original 

position (upside down if it was right side up, right side up if it 

was upside down)�

6�   Using parts from the factory kit, assemble the spring and sur-

rounding hardware according to the drawing of the desired 

configuration� (Note: All original parts may not be used when 

converting from close on increasing to open on increasing�)

7�   The entire bracket assembly or the outer angle may need 

to be turned upside down to accommodate the new spring 

height�

8�   Reconnect the supply, instrument, and output lines according 

to piping schematic supplied�

 

NOTE: The flow direction must be maintained through the 

positioner bodies when re-piping� (i�e�, the flow [supply or 

exhaust] moves from P1 to P2 and P3 to P4)�

Table 4. Extension Rods for Positioners to Close on  
Increasing Signal

Conversion to Split Range

Converting a standard positioner to a split range positioner (pneu-

matic input other than 3-15 psi or 6-30 psi), requires ordering the 

proper conversion kit from the factory� This kit will include a bias 

spring and bias spring cartridge� If required, it will also contain a 

new range spring and mounting spacers�

1�   For close on increasing positioner: Remove the cap on the top 

of the positioner along with the spring inside it�

 

For open on increasing positioner: Remove the mounting 

bracket holding the cap on the bottom of the positioner� Then 

remove this cap along with the spring inside it�

2�   Replace the cap and spring with the larger bias spring 

cartridge and bias spring found in the kit� Make sure the bias 

adjustment screw in the bias spring cartridge is snug against 

the bias spring and the spring is centered before tightening 

the spring cartridge�

3�   For open on increasing positioner: Remove the washer and 

jam nut from the adjusting screw in the bias spring cartridge� 

Reattach the mounting bracket upside down from its original 

position�

4�   Slide the thread spacer (brass bushing) over the adjusting 

screw and tighten the washer and jam nut against the thread 

spacer�

5�   If a range spring was sent with the kit, remove the existing 

range spring and replace it with the new one�

6�   Adjust the unit per the Adjustment Procedures�

Stroke

With Transmitter

Without Transmitter

4”

25-1402

25-8001

6”

25-8136

25-1093

8”

25-1402

25-1423

NOTE: Refer to Table 1 and Table 2 for bias and range spring part 

numbers

Table 5. Split Range Conversion Kits

Maintenance and Inspection

As with all precision equipment, it is necessary to periodically test 

the positioner to help ensure top performance� We recommend 

the following procedure once a year�

1�   Shut off supply pressure and bleed down at positioner� Note 

the settings of the variable orifices and remove them from 

the orifice assembly� Clean them thoroughly and reinstall 

using new o-rings, while being sure to install each orifice in 

the same hole from which it was removed (the orifice and 

block have matching numbers for this purpose)� Reset orifices 

to original settings� Turn on supply pressure�

2�   Apply a midrange signal to the positioner� Allow the control 

valve to become stationary at about 50 percent of the range� 

Close the cylinder block valves or move the MCV-3 handle 

to the manual position� The positioner is now isolated from 

the cylinder� Apply a ± 1/4 psig signal change� Observe the 

response in the output gauges� The output pressure should 

develop differential pressure equal to 20 percent of the power 

gas pressure� If the output pressure does not show immedi-

ate response, the positioner may have too much deadband� 

Reduce the deadband by turning the drum in the direction 

of decreasing numbers� If the pressures do not respond in 

the correct direction when reversing the instrument signal 

change, the unit has internal friction� Disassemble the unit 

and replace all rubber goods�

3�   Check the integrity of the balance valve seats by increasing 

the deadband by one full number� When the cylinder top and 

bottom gauges are equal to the power gas, the exhaust port 

should not exhibit any bleed gas� If either of these tests fail, 

then the positioner is not properly adjusted or the unit needs 

to be reassembled with new rubber goods�

4�   Soap test around all diaphragm interfaces, orifice  

assemblies, and vents� If any leaks are found around the 

diaphragms, refer to the assembly instructions for replace-

ment of all internal rubber parts�

5�   Observe the operation of the gauges� If any gauges are 

defective, replace them�

6�   Check range and bias� If necessary, readjust per Adjustment 

Procedures� Should problems arise or more information is 

required, call toll free

 (800) 323-8844 

for assistance�

Stroke

Close on 

Increasing

Open on  

Increasing

Reference  

Drawings

4 inch

25-6014

25-1464

Proportional:
35-0513
35-0313/A
35-0511
35-0511/A

6 inch

25-6014

25-1465

Proportional: 
35-0522
35-0529

8 inch

26-6014

25-1466

Tailrod:

Summary of Contents for Becker HPP-5

Page 1: ...GE Oil Gas GE Data Classification Public HPP 5 High Pressure Positioner Installation and Maintenance Manual ...

Page 2: ...ower pressure downstream or fuel gas system eliminating bleed gas completely This means significant savings because it limits expensive bleed gas and decreases the environmental impact of atmospheric hydrocar bons on diminishing natural resources Description The Becker HPP 5 positioner is used in control valve assemblies with a controller or I P transducer and double acting actuator to provide acc...

Page 3: ...t GE for additional information Retrofit Compatibility Excellent performance is achieved by pairing the HPP 5 positioner with genuine GE control valve actuators Should you already have existing control valve actuator s in service the addition of a Model HPP 5 positioner can improve performance and eliminate atmospheric bleed emissions if using the BPS Bleed to Pressure System feature Some Compatib...

Page 4: ...the supply and discharge pressures In steady state the forces imposed on the balance beam by the input signal diaphragm and the range spring are equal therefore the top and bottom balance valves in the positioner are at or near their closed positions The cylinder top and bottom pressures are both equal to power gas pressure and the control valve is stationary An increase in the input signal pressu...

Page 5: ...greater accuracy from a smaller deadband and minimal bleed gas Your HPP 5 positioner will come factory adjusted for your particular application The use of the adjustment procedures will be necessary upon installation of a rubber goods replacement kit or any other disassembly or reassembly of the positioner Adjustment Procedure The sensitivity adjustment drum on one end of the positioner determines...

Page 6: ...ual position If the actuator is not equipped with an MCV 3 manual control valve use the block valve 1 4 inch ball valve installed between the positioner output and the actuator Once the actuator has reached its desired position based on the mid signal applied to the instrument port Step 5 place the block valve in the closed position 7 Final Sensitivity Drum Adjustment for positioner s bleeding to ...

Page 7: ... the start of the valve travel continue increasing the instrument signal until the valve stroke is completed and the cylinder output gauges show full power gas pressure differential 2 Be sure not to overshoot at this point as any signal level above this upper limit will show the same reading on the cylinder gauges This point is the highest end of the range and the value may not necessarily be equa...

Page 8: ...ge spring was sent with the kit remove the existing range spring and replace it with the new one 6 Adjust the unit per the Adjustment Procedures Stroke With Transmitter Without Transmitter 4 25 1402 25 8001 6 25 8136 25 1093 8 25 1402 25 1423 NOTE Refer to Table 1 and Table 2 for bias and range spring part numbers Table 5 Split Range Conversion Kits Maintenance and Inspection As with all precision...

Page 9: ...des adequate preventative maintenance Parts Ordering The following is provided to allow the ordering of replacement parts Please specify the Becker instrument serial number when ordering parts this can be found on the stainless steel tab attached to the pilot by the 7 16 hex head cap screws If the instrument was supplied as a complete valve regulator package the stainless tag attached to the actua...

Page 10: ...ORIFICE ASSY 7 1 35 1519 SEAT COVER 8 1 98 3056 1 2 20 JAM NUT SS 9 1 25 1124 BEAM ADJUSTING DRUM 10 1 25 1213 DIAPHRAGM 11 6 95 2609 O RING 010 12 1 25 1065 1 2 20 SPECIAL NUT ALUM 13 1 98 2950 1 8 ROLL PIN SS 14 4 98 3144 8 32 x 1 SHCS 15 1 25 1086 BEAM BLOCK 16 1 35 1506 OUTSIDE PISTON 17 4 98 3089 3 16 x 1 2 ROLL PIN SS 18 4 98 3180 1 4 20 x 2 1 2 HHCS 19 1 35 1013 GAUGE MANIFOLD 20 6 95 2615 ...

Page 11: ...E ON INCREASING SIGNAL 21 ROD CLEVIS HOUSING ASSY INSTRUMENT GAUGE HPP 09 13 93 NTS TS DRAWING NUMBER APPR0VED BY B DWG SIZE REVISED BY DB REVISED DATE 2 12 04 DRAWN BY DATE SCALE 30 1208 ITEM QTY PART DESCRIPTION 1 1 95 2643 O RING 24 2 1 25 1095 SPRING RETAINER 3 1 95 2644 O RING 30 4 1 25 1289 SHORT TUBING 5 1 TUBE HOLDER 6 1 25 1310 ACCESS COVER 7 1 25 1131 BOTTOM COVER 8 1 25 1230 8 STK HPP 2...

Page 12: ... POSITION 2 THE ORIFICE ASSEMBLY IS CALIBRATED TO ASSURE EQUAL RATE OF FLOW FROM BOTH ORIFICES 1 REGULATOR TO CLOSE ON INCREASING SIGNAL N0TES HPP INSTRUMENT GAUGE BECKER Precision Equipment DRESSER INC ITEM QTY PART DESCRIPTION 1 1 50 1230 FLANGE 2 1 25 1310 ACESS COVER 3 1 98 3171 8 32 X 1 4 RHMS 4 1 25 1149 RANGE SPRING LOCK 5 1 RANGE SPRING 6 1 95 2643 O RING 024 7 1 25 1134 LONG TUBING 8 1 95...

Page 13: ...b Insert the balanced valve assembly C with the stem facing downward 1c Insert the second balanced valve assembly C with the stem facing upward 1d Insert the second strainer D1 and spacer D assembly 1e Insert the top seat rubber facing downward E 1f Secure the seat assembly to the pilot body block F with two washers C1 seat cover B and two flat head machine screws A NOTE Install washers to compens...

Page 14: ...h four 1 4 20 x 2 1 2 inch stainless steel hex head cap screws J by lining it up with roll pins M in pilot body block F Step 4 Test the pilot body block F for leakage Using 1 4 inch NPT plugs plug the orifice block ports K and the gauge manifold N 4a Connect a pressure source to the orifice block K as shown Use approximately 100 psig for testing 4b With the pressure source activated soap test arou...

Page 15: ...ly with two 8 32 x 1 inch SHCS T Step 6 Slide 012 O ring CC onto the outside piston I and into the groove near its base 6a Install one washer Z onto the shaft of the outside piston I with the grooves facing upward 6b Install the convoluted diaphragm AA onto the shaft of the outside piston as shown 6c Install one washer Z onto the shaft of the outside piston with the grooves facing downward 6d Tigh...

Page 16: ...ith six 1 4 20 x 3 4 inch hex head cap screws DD Align holes in the beam spacer JJ and in the beam adjusting drum KK as shown Step 10 Screw the beam drum KK on to the piston F until it stops Then turn the beam drum the opposite direction until the slot in the beam drum aligns with that of the beam spacer Step 11 Place the beam NN inside the beam block LL in the direction shown and then place the 1...

Page 17: ...Becker HPP 5 High Pressure Positioner Maintenance and Operation Manual 16 2014 General Electric Company All rights reserved ...

Page 18: ... ...

Reviews: