Part Number 89604-01 

Rev. F (08/07) 

Bently Nevada™ Asset Condition Monitoring 

 

Operation Manual 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

3300/03 System Monitor 

 
 

Summary of Contents for Bently Nevada 3300/03

Page 1: ...Part Number 89604 01 Rev F 08 07 Bently Nevada Asset Condition Monitoring Operation Manual 3300 03 System Monitor ...

Page 2: ...Company in the United States and other countries Dynamic Data Manager Keyphasor Proximitor and Transient Data Manager Contact Information The following ways of contacting Bently Nevada are provided for those times when you cannot contact your local representative Mailing Address 1631 Bently Parkway South Minden Nevada USA 89423 USA Telephone 1 775 782 3611 1 800 227 5514 Fax 1 775 215 2873 Interne...

Page 3: ...ta Manager User Guide Part Number 79206 01 Product Disposal Statement Customers and third parties who are not member states of the European Union who are in control of the product at the end of its life or at the end of its use are solely responsible for the proper disposal of the product No person firm corporation association or agency that is in control of product shall dispose of it in a manner...

Page 4: ...IONS 2 3 ASSEMBLY DISASSEMBLY PROCEDURE 4 4 DATA INTERFACE REMOVAL 5 5 FRONT PANEL INSTALLATION REMOVAL 6 6 MONITOR OPTIONS 7 7 PROGRAMMABLE OPTIONS 8 8 OK RELAY CONFIGURATION 10 9 OPERATIONAL TESTING 11 10 Keyphasor OPERATIONAL TEST 12 11 RECOMMENDED SPARE PARTS 17 12 SPECIFICATIONS 18 ...

Page 5: ...Section 1 SYSTEM MONITOR 1 1 SYSTEM MONITOR ...

Page 6: ...sing the RESET switch on the front panel will cause a System Reset TRIP MULTIPLY The most common use of this function is to prevent unwanted monitor alarms during certain conditions of machine operation In the operation of some machinery it is impossible to avoid some periods of higher than normal vibration Examples include startup and coastdown especially if the operating speed is above rotor sys...

Page 7: ...evada data interfaces Dynamic Data Manager DDM Transient Data Manager TDM Serial Data Interface SDI Dynamic Data Interface DDI and Transient Data Interface TDI The SDI and DDI are options of the System Monitor and reside within the System Monitor slot of the rack DDM TDM and TDI are external data interfaces which gain access to the rack through the PIM There is an external version of the SDI and D...

Page 8: ...nitor to a rack set the options as described in section 7 then do steps 1 through 4 in reverse order The only tool you need is a screwdriver 1 Loosen the screws on the front panel and pull the System Monitor out from the rack 2 Remove the side cover by pinching the protruding tip on each of the 4 standoffs CAUTION Improper rack operation may occur Power down rack when intalling or removing a monit...

Page 9: ...rface boards to access the option jumpers located on the System Monitor 1 Remove the Dynamic Data Interface board by pinching the protruding tip on each of the 4 standoffs and gently prying the Dynamic Data Interface board away from the Serial Data Interface board 2 Remove the Serial Data Interface board by gently prying it away from the two mating connectors and 5 standoffs on the System Monitor ...

Page 10: ...3300 03 System Monitor Operation Manual 6 5 FRONT PANEL INSTALLATION REMOVAL ...

Page 11: ...Section 6 MONITOR OPTIONS 7 6 MONITOR OPTIONS SYSTEM MONITOR PART NUMBER DATA INTERFACE AGENCY APPROVAL 3300 03 AXX BXX 01 None or DDM TDM 02 SDI 03 DDI 00 NOT REQUIRED 01 CSA 02 BASEEFA 03 FM ...

Page 12: ...AMMABLE OPTIONS The System Monitor has jumper programmable options for selecting the transducer voltage and the data interface Change these options by removing and installing jumpers on the circuit board Jumper Locations W6 is for future use or is N A ...

Page 13: ...Interface External Serial Data Interface Dynamic Data Interface or Transient Data Interface W3A W3B W5A W5B W5C W5D W5E W5F W5G W5H W8D W8E W3C W3D W4A W4B W4C W4D W4E W4F W4G W4H W7 W8D W8E W3C W3D W5A W5B W5C W5D W5F W5G W5H W8A W8B W8C As shipped from factory for a 3300 03 01 System Monitor As shipped from factory for a 3300 03 02 or 3300 03 03 System Monitor VT Options To set the VT option acc...

Page 14: ...3300 03 System Monitor Operation Manual 10 8 OK RELAY CONFIGURATION The following diagram shows the functional concept of the OK Relay For more detail refer to schematics see Section 13 ...

Page 15: ...he system monitor is functional apply power at the Power Input Module The LED indicating that the supplies are OK should turn on after 2 seconds CAUTION Improper operation may occur Set VT Jumpers in the System Monitor and Power Supply to the same voltage 18 VDC or 24 VDC ...

Page 16: ...ducer under test should be running during this procedure If the machine is not turning simulate the operating machinery input at approximately the same frequency as the shaft rotating speed with a function generator or TK3 test kit 1 Measure Keyphasor 1 power at the KPH1PWR terminal on the power input module PIM If VT is 24 V the voltage should read between 23 2 and 24 2 Vdc If VT is 18 V the volt...

Page 17: ...inal The voltage should be within the tolerance specified in step 1 3 Loosen the screws that hold the power input module to the rack and pull the power input module out from the rear of the rack Examine the cabling and connections between the module and the rack WARNING High voltage present Contact could cause shock burns or death ...

Page 18: ...KPH1PWR terminal Disconnect the wire at the VT 18 or 24 Vdc terminal of the Proximitor Measure the voltage at the wire The voltage should be within the tolerances specified in step 1 5 Check the Keyphasor output signal at the System Monitor front panelKO 1coaxialconnector ...

Page 19: ...Section 10 Keyphasor OPERATIONAL TEST 15 6 Check Keyphasor signal at the power input module KPH1 terminal 7 Disconnect wire at the power input module KPH1 terminal and check wire for Keyphasor signal ...

Page 20: ...3300 03 System Monitor Operation Manual 16 8 Reconnect wire to the power input module KPH1 terminal 9 Disconnect wire at the Proximitor output terminal and check signal at the output terminal ...

Page 21: ...7890 01 1 Spare Jumpers 100 pieces 88706 01 TO ORDER REPLACEMENT PARTS SPECIFY THE COMPLETE CATALOG NUMBER 3300 03 AXX BXX AND THE REPLACEMENT PART NUMBER If the monitor has been modified specify the modification number on the parts order You must set the programmable options Section 7 lists how the options are set at the factory If in doubt about the part number call your Bently Nevada Corporatio...

Page 22: ...V Option 23 75 0 45 Vdc VT 18V Option 17 9 0 4 Vdc SYSTEM POWER UP INHIBIT Delay 2 1 1 seconds CONTACT CLOSURE RATING Contacts for System Reset Inhibit OK and Trip Multiply Active 100 Ω Keyphasor Output Voltage 24V Option 23 2 V to 24 2 V Short circuit Protected 18V Option 17 4 V to 18 3 V Short circuit Protected ENVIRONMENTAL Temperature Operating 32 F to 149 F 0 C to 65 C Storage 40 F to 185 F 4...