EX
WARNING
EXPLOSION HAZARD! To avoid building up an explosive atmosphere
when using flammable liquids, make sure that the room ventilation
meets the requirements specified in BioProcess LPLC or MPLC
columns Installation Guide.
EX
WARNING
EXPLOSION HAZARD! When a BioProcess LPLC or MPLC column is
operated or maintained in a potentially explosive atmosphere, it
MUST be properly grounded to avoid static discharge.
EX
WARNING
EXPLOSION HAZARD!
Avoid discharge from static electricity when working in potentially
explosive atmospheres by:
•
Always using suitable clothing.
•
Making sure that the entire system including the frame, piping
and column are connected to the plant grounding network.
•
Use only non-sparking tools rated for EX-work.
EX
WARNING
EXPLOSION HAZARD!
Only tools and equipment for use in potentially explosive atmo-
spheres should be used for operation and maintenance.
EX
WARNING
Do not clean polymerics parts with a dry cloth or similar.
LPLC and MPLC Columns Operating Instructions 28-9638-66 AC
19
2 Safety instructions
2.1 Safety precautions