Troubleshooting
6-39
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0311546
Error
Administration
module
unsubscribe
sync event
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0309500
Warning
Setup.xml does
not exist
Setup configuration loading
failure. Please restart the
system. If this issue persists,
please call for service.
check if configuration file
crashed
0309501
Warning
Setup
Default.xml
does not exist
Default Setup configuration
loading failure. Please restart
the system. If this issue
persists, please call for service.
check if default
configuration file crashed
0309502
Warning
ConnectivitySet
up.xml is not
exist
Connectivity configuration
loading failure. Please restart
the system. If this issue
persists, please call for service.
check if Dicom
configuration file crashed
0309503
Warning
ConnectivitySet
upDefault.xml is
not exist
Default connectivity
configuration loading failure.
Please restart the system. If
this issue persists, please call
for service.
check if default Dicom
configuration file crashed
0309504
Warning
Administration.
xml is not exist
Administration configuration
loading failure. Please restart
the system. If this issue
persists, please call for service.
check if admin
configuration file crashed
0309505
Warning
ExposureContro
l.xml is not exist
Exposure control configuration
file loading failure. Please
restart the system. If this issue
persists, please call for service.
check if exposure
configuration file crashed
0309506
Warning
SwBuild.xml is
not exist
Software version access
failure. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0309507
Warning
SerialNumber.x
ml is not exist
Serial number access failure.
Please restart the system. If
this issue persists, please call
for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...