![background image](http://html.mh-extra.com/html/ge/carescape-b105/carescape-b105_supplement-manual_128558029.webp)
ApexPro and CARESCAPE Telemetry
Disinfectant
Manufacturer
Ingredients
Type of disinfectant
PDI Super Sani-Cloth
Germicidal
Disposable Wipes*
Professional
Disposables Inc.
●
n-Alkyl (68% C12, 32% C14)
dimethyl ethylbenzyl ammonium
chlorides 0.25%
●
n-Alkyl (60% C14, 30% C16, 5%
C12, 5% C18) dimethyl benzyl
ammonium chlorides 0.25%
●
Isopropanol 55.00%
Quaternary
ammonium and
alcohol
PDI Easy Screen
Cleaning Wipes*
Professional
Disposables Inc.
●
Isopropyl alcohol 70%
Alcohol
Oxivir TB Wipes*
Sealed Air
●
Hydrogen peroxide 0.1 to 1.5%
●
Benzyl alcohol 1 to 5%
Accelerated
hydrogen peroxide
Ethanol 70 to 96%
Generic
●
Ethanol
Alcohol
Isopropyl alcohol
max. 60% by weight
Generic
●
Isopropyl alcohol
Alcohol
Bleach [dilute to max.
of 0.65%]
Generic
●
Sodium hypochlorite
Chlorine
* Available in the United States.
Disinfecting external surfaces (telemetry monitoring
devices only)
The decision to disinfect must be made per the institution’s requirements with an
awareness of the effect on the integrity of the device. Do not use excessive drying
techniques (e.g., oven, forced heat, sun drying).
1.
Remove all cables and the batteries.
2.
Close the battery door.
3.
Dampen a soft lint-free cloth, or use disinfectant wipes with one of the permitted
disinfectants.
4.
Wring excess liquid from the cloth and wipe the exterior surface. Pay special
attention to hard-to-clean areas, like grooves and crevices. Allow disinfecting
solution to remain on device per disinfectant instructions or per hospital
guidelines.
●
Any contact of disinfecting solutions with metal parts may cause corrosion.
●
Do not damage or bend connector pins when cleaning or drying.
●
Do not let fluid pool around connection pins. If this happens, blot dry with
a cotton swab or soft cloth.
5.
Wipe off the disinfecting solution with a clean, lightly moistened cloth.
6.
Dry thoroughly with a dry, lint-free cloth and let air dry for at least 30 minutes.
Drying times may vary based on the environmental conditions.
7.
Install the batteries and close the battery door.
8.
Reconnect all cables.
2110145-001
Cleaning and disinfecting
29
Summary of Contents for CARESCAPE B105
Page 8: ...8 Cleaning and disinfecting 2110145 001 ...
Page 30: ...ApexPro and CARESCAPE Telemetry 30 Cleaning and disinfecting 2110145 001 ...
Page 48: ...CARESCAPE B450 Monitor 48 Cleaning and disinfecting 2110145 001 ...
Page 54: ...CARESCAPE B650 Monitor 54 Cleaning and disinfecting 2110145 001 ...
Page 71: ...content ...