PRN 50M+
Cleaning the print head
This procedure explains how to clean the thermal print head. From heavy usage,
debris accumulates on the thermal print head and distorts the printed image. It is
recommended that you perform this procedure when necessary, depending on usage.
A nonabrasive material/cloth and isopropyl alcohol are all that are necessary to
perform this procedure.
1.
Turn off the power to the device.
2.
Disconnect the device from the power supply.
3.
Open the writer door to expose the print head.
4.
Remove paper roll.
5.
Locate the print head. A flashlight may help illuminate the print head for closer
examination.
6.
Wipe the print head with alcohol and a nonabrasive material/cotton swab in a
side to side motion. Continue wiping until the cloth/swab wipes clean.
7.
Wipe paper drive roller clean of any bits of paper and debris with alcohol and a
nonabrasive material.
Disinfection
Disinfection notices
NOTICE
COLOR SHIFT. Some plastics may experience a slight change
in color over time due to effects of disinfecting agents. This
effect is cosmetic and not associated with product safety or
normal function.
NOTICE
SPOTTING, STAINING, CORROSION. Metal components
including fasteners may corrode or become stained or
spotted in appearance from continuous exposure to chlorine
compounds like bleach. These effects are not associated with
product safety or normal function.
Disinfection points to note
●
Always clean before disinfecting.
●
Always dilute cleaning and disinfectant agents according to their manufacturer’s
instructions. Always consider your hospital guidelines as well.
●
Use only the permitted substances.
●
Your hospital guidelines permitting, all cleaning and disinfection activities can
be carried out at the bedside.
●
Before drying a surface after wiping it, wait the minimum time required according
to the substance manufacturer’s instructions.
●
Visually ensure that no substance residue remains on the device.
Permitted disinfectants
About the following table:
2110145-001
Cleaning and disinfecting
67
Summary of Contents for CARESCAPE B105
Page 8: ...8 Cleaning and disinfecting 2110145 001 ...
Page 30: ...ApexPro and CARESCAPE Telemetry 30 Cleaning and disinfecting 2110145 001 ...
Page 48: ...CARESCAPE B450 Monitor 48 Cleaning and disinfecting 2110145 001 ...
Page 54: ...CARESCAPE B650 Monitor 54 Cleaning and disinfecting 2110145 001 ...
Page 71: ...content ...