Stopping High Flow 
Oxygen Therapy to  
Begin Invasive or 
Noninvasive Ventilation

1.  Safeguard the patient. 

2. Select Current Mode.

3.  Select a mode from the list of modes.

4.  Adjust the ventilation mode settings as needed.

5. Select Confirm from the ventilation mode window.

6. Select PAUSE O2 THERAPY from WARNING window.

7.  Confirm or complete a SYSTEM CHECK for the dual-limb  
  circuit type.

Note:

 If using a single-limb circuit, remove the circuit and  

connect a dual-limb circuit to the ventilator.

8. Select START VENTILATION.

Note: 

A Neonatal Flow Sensor Calibration should be performed  

upon returning to Mechanical Ventilation, if the Neonatal Flow  
Sensor was not previously calibrated manually or automatically  
during a system check using a dual-limb circuit. 

2

5

6

7

8

3

4

Summary of Contents for CARESCAPE R860

Page 1: ...gehealthcare com CARESCAPE R860 Ventilator O2 Therapy Quick Guide...

Page 2: ...dures or maintenance methods for any of the products referenced Always refer to the official written materials labeling provided with the product for specifications operating procedures and maintenanc...

Page 3: ...ow Oxygen Therapy After System Start up 8 Starting O2 Therapy from Invasive or Noninvasive Ventilation 10 Navigating O2 Therapy Present View 11 Stopping High Flow Oxygen Therapy 11 Stopping High Flow...

Page 4: ...nt interface High flow oxygen therapy offers a number of physiological benefits such as pharyngeal dead space washout work of breathing reduction PEEP effect constant FiO2 and improvement of mucocilia...

Page 5: ...to the humidifier base 2 Connect Circuit Dry Line to humidification chamber and Inspiratory Safety Guard 3 Connect the inspiratory limb of the breathing circuit to the humidification chamber and the e...

Page 6: ...te An inspiratory safety guard must be installed on the inspiratory port and any accessories that will be used Do not attach any accessories that prevent attachment of the patient wye to the occlusion...

Page 7: ...e installed on the inspiratory port and any accessories that will be used 4 Select Start 5 Select the information icon to see the System Check Details menu a As the system check is running the results...

Page 8: ...to the inspiratory port 2 Select NEW PATIENT to select the appropriate patient type and add any additional demographic data 3 Select Current Mode 4 Select O2 Therapy from the list of modes 5 Set the F...

Page 9: ...ing High Flow Oxygen Therapy After System Start up cont 9 Complete a SYSTEM CHECK for the appropriate circuit type 10 Select START O2 THERAPY Note The system delivers Oxygen and the O2 Therapy Present...

Page 10: ...nt Mode 3 Select O2 Therapy from the list of Modes 4 Set the FiO2 5 Set the Flow 6 Select Confirm to confirm O2 Therapy settings 7 Select Pause Ventilationfromthe O2Therapy confirmation WARNING window...

Page 11: ...graph is not patient airway pressure The pressure represents the resistance to flow provided by the cannula and the applied patient interface and may indicate an occlusion to the patient interface dur...

Page 12: ...E O2 THERAPY from WARNING window 7 Confirm or complete a SYSTEM CHECK for the dual limb circuit type Note If using a single limb circuit remove the circuit and connect a dual limb circuit to the venti...

Page 13: ...out notice or obligation To determine whether individual features are standard or optional consult with your GE Healthcare sales representative for the most current information GE the GE Monogram and...

Reviews: