12
Stopping High Flow
Oxygen Therapy to
Begin Invasive or
Noninvasive Ventilation
1. Safeguard the patient.
2. Select Current Mode.
3. Select a mode from the list of modes.
4. Adjust the ventilation mode settings as needed.
5. Select Confirm from the ventilation mode window.
6. Select PAUSE O2 THERAPY from WARNING window.
7. Confirm or complete a SYSTEM CHECK for the dual-limb
circuit type.
Note:
If using a single-limb circuit, remove the circuit and
connect a dual-limb circuit to the ventilator.
8. Select START VENTILATION.
Note:
A Neonatal Flow Sensor Calibration should be performed
upon returning to Mechanical Ventilation, if the Neonatal Flow
Sensor was not previously calibrated manually or automatically
during a system check using a dual-limb circuit.
2
5
6
7
8
3
4