6
1
High Flow Oxygen Therapy System Check
Be sure that the ventilator is prepared for mechanical ventilation
in an emergency scenario. It is recommended that a full
(dual-limb) system check be completed prior to use of the
O2 Therapy feature with a single-limb configuration.
To ensure the proper function of the system, it is highly
recommended to complete a System Check before use and
between patients. Failure to complete a System Check may
result in inaccurate delivery and monitoring.
Note:
O2 Therapy is allowed even if the safety relief valve test
fails during the system check.
Dual-limb Circuit
1. From Standby, select SYSTEM CHECK.
Note:
The Run System Check menu shows.
2. Select the dual-limb circuit icon.
3. Attach the dual-limb breathing circuit and all accessories
that will be used during O2 Therapy.
Note:
An inspiratory safety guard must be installed on the
inspiratory port and any accessories that will be used. Do not
attach any accessories that prevent attachment of the patient
wye to the occlusion port.
4. Occlude the patient wye using the occlusion port.
Note:
The occlusion port sits below the exhalation valve
assembly.
5. Select Start.
6. Select the information icon to see the System Check
Details menu.
a. As the system check is running, the results of each
check will be displayed.
Note:
Follow all on-screen system check instructions. A green
check mark will display if the check passes. If the check fails,
a red X is shown, and a Help icon displays next to the failed
check to provide possible causes and help for troubleshooting.
The System Check results show the dual-limb circuit icon, the
result of the check, and the date and time of the check.
7. Select START O2 THERAPY to begin delivery of Oxygen.
8. Confirm Oxygen is flowing through the patient circuit
before connecting to the patient.
1
2
6a
5
7
6