background image

6

1

High Flow Oxygen Therapy System Check 

Be sure that the ventilator is prepared for mechanical ventilation 
in an emergency scenario. It is recommended that a full  
(dual-limb) system check be completed prior to use of the  
O2 Therapy feature with a single-limb configuration.

To ensure the proper function of the system, it is highly  
recommended to complete a System Check before use and 
between patients. Failure to complete a System Check may 
result in inaccurate delivery and monitoring.

Note: 

O2 Therapy is allowed even if the safety relief valve test 

fails during the system check.

Dual-limb Circuit

1.  From Standby, select SYSTEM CHECK.

Note:

 The Run System Check menu shows.

2.  Select the dual-limb circuit icon.

3.  Attach the dual-limb breathing circuit and all accessories  
  that will be used during O2 Therapy.

Note: 

An inspiratory safety guard must be installed on the 

inspiratory port and any accessories that will be used. Do not 
attach any accessories that prevent attachment of the patient 
wye to the occlusion port.

4.  Occlude the patient wye using the occlusion port.

Note:

 The occlusion port sits below the exhalation valve  

assembly.

5. Select Start.

6.  Select the information icon to see the System Check  
  Details menu.

  a.  As the system check is running, the results of each  
    check will be displayed.

Note:

 Follow all on-screen system check instructions. A green 

check mark will display if the check passes. If the check fails, 
a red X is shown, and a Help icon displays next to the failed 
check to provide possible causes and help for troubleshooting. 
The System Check results show the dual-limb circuit icon, the 
result of the check, and the date and time of the check.

7. Select START O2 THERAPY to begin delivery of Oxygen.

8.  Confirm Oxygen is flowing through the patient circuit  
  before connecting to the patient.

1

2

6a

5

7

6

Summary of Contents for CARESCAPE R860

Page 1: ...gehealthcare com CARESCAPE R860 Ventilator O2 Therapy Quick Guide...

Page 2: ...dures or maintenance methods for any of the products referenced Always refer to the official written materials labeling provided with the product for specifications operating procedures and maintenanc...

Page 3: ...ow Oxygen Therapy After System Start up 8 Starting O2 Therapy from Invasive or Noninvasive Ventilation 10 Navigating O2 Therapy Present View 11 Stopping High Flow Oxygen Therapy 11 Stopping High Flow...

Page 4: ...nt interface High flow oxygen therapy offers a number of physiological benefits such as pharyngeal dead space washout work of breathing reduction PEEP effect constant FiO2 and improvement of mucocilia...

Page 5: ...to the humidifier base 2 Connect Circuit Dry Line to humidification chamber and Inspiratory Safety Guard 3 Connect the inspiratory limb of the breathing circuit to the humidification chamber and the e...

Page 6: ...te An inspiratory safety guard must be installed on the inspiratory port and any accessories that will be used Do not attach any accessories that prevent attachment of the patient wye to the occlusion...

Page 7: ...e installed on the inspiratory port and any accessories that will be used 4 Select Start 5 Select the information icon to see the System Check Details menu a As the system check is running the results...

Page 8: ...to the inspiratory port 2 Select NEW PATIENT to select the appropriate patient type and add any additional demographic data 3 Select Current Mode 4 Select O2 Therapy from the list of modes 5 Set the F...

Page 9: ...ing High Flow Oxygen Therapy After System Start up cont 9 Complete a SYSTEM CHECK for the appropriate circuit type 10 Select START O2 THERAPY Note The system delivers Oxygen and the O2 Therapy Present...

Page 10: ...nt Mode 3 Select O2 Therapy from the list of Modes 4 Set the FiO2 5 Set the Flow 6 Select Confirm to confirm O2 Therapy settings 7 Select Pause Ventilationfromthe O2Therapy confirmation WARNING window...

Page 11: ...graph is not patient airway pressure The pressure represents the resistance to flow provided by the cannula and the applied patient interface and may indicate an occlusion to the patient interface dur...

Page 12: ...E O2 THERAPY from WARNING window 7 Confirm or complete a SYSTEM CHECK for the dual limb circuit type Note If using a single limb circuit remove the circuit and connect a dual limb circuit to the venti...

Page 13: ...out notice or obligation To determine whether individual features are standard or optional consult with your GE Healthcare sales representative for the most current information GE the GE Monogram and...

Reviews: