8
Starting High Flow Oxygen Therapy After
System Start-up
1. Confirm the appropriate patient circuit and humidified
patient interface is connected to the inspiratory port.
2. Select NEW PATIENT to select the appropriate patient
type and add any additional demographic data.
3. Select Current Mode.
4. Select O2 Therapy from the list of modes.
5. Set the FiO2%.
6. Set the Flow.
7. Select Confirm to confirm O2 Therapy settings.
8. Select Confirm from the O2 Therapy confirmation
WARNING window.
2
3
7
4
8
5
6