System safety
Preparing for use
Read each component’s User’s Reference manual and
understand the following before using this system:
•
All system connections.
•
All warnings and cautions.
•
How to use each system component.
•
How to test each system component.
Before using the system:
•
Complete all of the tests in the "
"
section.
•
Test all other system components.
•
If a test fails, do not use the equipment. Have an
authorized service representative repair the
equipment.
European, international, and national standards require
the following monitoring be used with this system:
•
Exhaled volume monitoring.
•
O2 monitoring.
•
CO2 monitoring.
•
Anesthetic agent monitoring be used when anesthetic
vaporizers are in use.
Single-use products are not designed or validated to be
reused. Reuse may cause a risk of cross-contamination,
affect the measurement accuracy, system performance,
or cause a malfunction as a result of the product being
physically damaged due to cleaning, disinfection, re-
sterilization, or reuse.
Follow hospital procedures for the prevention and
treatment of malignant hyperthermia for patients sensitive
to inhalation anesthetic agents.
WARNING
•
•
•
•
1 Introduction
2076152-001
1-19
Summary of Contents for Carestation 620
Page 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Page 12: ...Carestation 620 650 650c A1 x 2076152 001...
Page 58: ...2 24 2076152 001...
Page 98: ...3 40 2076152 001...
Page 102: ...4 4 2076152 001...
Page 128: ...6 12 2076152 001...
Page 150: ...7 22 2076152 001...
Page 182: ...8 32 2076152 001...
Page 196: ...9 14 2076152 001...
Page 246: ...11 44 2076152 001...
Page 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...