3.
Select
Close
.
Setting alarm volume
Set alarm volume loud enough to be heard over
surrounding noise. If the volume is not loud enough, the
user may not hear the alarm.
1.
Select
Alarm Setup
.
2.
Select the
Configure
tab.
3.
Set
Alarm Volume
to the desired value.
The alarm volume range is 1 to 5.
4.
Select
Close
.
Setting apnea delay
Use the
Apnea Delay
setting to set the desired apnea time delay.
The apnea time delay is the amount of time that can pass without the
system detecting a measured breath before the apnea alarm occurs.
1.
Select
Alarm Setup
.
2.
Select the
Configure
tab.
3.
Set
Apnea Delay
to the desired time.
•
The apnea time delay range is 10 to 30 seconds.
•
The apnea time settings are in 1 second increments.
4.
Select
Close
.
Silencing leak audio alarms
Use the
Leak Audio
setting to silence audio alarms of small leaks.
1.
Select
Alarm Setup
.
2.
Select the
Configure
tab.
3.
Set
Leak Audio
to
Off
.
The audio alarms for small leaks are silenced.
4.
To turn audio alarms back on, set
Leak Audio
to
On
.
5.
Select
Close
.
WARNING
3 Operation
2076152-001
3-23
Summary of Contents for Carestation 620
Page 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Page 12: ...Carestation 620 650 650c A1 x 2076152 001...
Page 58: ...2 24 2076152 001...
Page 98: ...3 40 2076152 001...
Page 102: ...4 4 2076152 001...
Page 128: ...6 12 2076152 001...
Page 150: ...7 22 2076152 001...
Page 182: ...8 32 2076152 001...
Page 196: ...9 14 2076152 001...
Page 246: ...11 44 2076152 001...
Page 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...