C-RS Ultrasonic Flow Transducer Installation Guide
21
6.1 The Double-Traverse Method - GCF (cont.)
5.
Make two marks on the line at the top of the pipe. The distance between these two marks must equal the calculated
transducer spacing
S
.
6.
From each of the marks, measure around the circumference of the pipe in the same direction a distance equal to 1/4
of the pipe circumference. Make a crossmark at each point.
Line
Top of Pipe
Side View
Spacing
Mark
Mark
Side View
Spacing
Crossmarks
Side View