C-RS Ultrasonic Flow Transducer Installation Guide
1
1. Introduction
The
Model C-RS
clamp-on ultrasonic flow transducer for liquids is used exclusively with the GE line of ultrasonic
flowmeters. These transducers measure the flow rate of sonically-conductive liquids through pipes having diameters
from 2 in. (5 cm) to over 300 in. (7.6 m). Such measurements are typically independent of the pipe material.
2. Transducer Construction
C-RS transducers may be used at process temperatures from -40 to 302
o
F (-40 to 150
o
C). Each C-RS transducer
assembly consists of the following components (see Figure 1):
•
A stainless steel adapter with 3/4” NPT male thread for attaching a junction box
•
A transducer that consists of a piezoelectric element mounted on a wedge and wired to the BNC connector
•
A BNC style connector for use in connecting the transducer to the flowmeter.
Note:
The process temperature range specified above is for ATEX certified designs. Consult the factory for higher
temperatures.
Figure 1: General C-RS Transducer Assembly
BNC Connector
Transducer Body
Adapter
Summary of Contents for CModel C-RS
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 52: ...44 C RS Ultrasonic Flow Transducer Installation Guide no content intended for this page ...
Page 56: ......
Page 57: ......