4-44
250 Series Maternal/Fetal Monitor
Revision D
2020551-001
Maintenance:
NIBP Calibration and Testing
All NIBP calibration procedures are accessed from the
NIBP Calibration
screen. To
access the NIBP Calibration screen:
1. Navigate to
Install Screen Options 1.
2. Select
Tests
.
3. Select
NIBP Cal.
The sequence of procedures for calibration of NIBP starts with Calibration
Verification. If Calibration Verification shows the NIBP transducers to be out of
calibration then calibration should be performed. Only after NIBP transducers have
been shown to be in calibration should any other test be performed. Both
Overpressure Detection and System Leakage should always be performed.
Calibration Verification
1. Select
Verify
from the
Mode:
option.
2. The monitor will inflate the cuff to approximately 200 mmHg.
3. If either PT1 or PT2 pressure, from the external manometer, are
> = 2 mmHg (0.3 kPa) different from the monitor indicated pressure then
perform Calibrate Transducers procedure.
4. To stop Calibration Verification select
Done,
which appears after the
Verify
softkey has been pressed.
5. The monitor will vent pressure to atmosphere and re-zero the transducers.
Calibrate Transducers
1. Select
Calibrate
from the
Mode:
option.
2. The monitor will inflate the cuff to approximately 200 mmHg (26.7 kPa).
3. Once the pressure has stabilized, enter the pressure from the external
manometer in
External:
4. Best accuracy is achieved if the system is given a short time to settle. Best
accuracy is achieved if the system pressure is at or near
200 mmHg (26.7 kPa).
5. Additional pressure can be inserted into the system using a hand bulb.
6. The monitor will vent pressure to atmosphere and re-zero the transducers.
7. Perform a Calibration Verification and repeat Calibration if new calibration
factors still show the transducers out of calibration
8. Commit the new calibration factors by selecting
Store Calibration.
NOTE
:
Between entering the external pressure and committing the new
calibration factors the
Exit
menu item will display as
Exit – No Store
to
indicate current calibration factors will be lost if Service Mode is
exited prior to selecting
Store Calibration.
NOTE
:
The menu item
Store Calibration
will only display after Calibration
Verification has been performed during the Calibration procedure.
Summary of Contents for Corometrics 250 Series
Page 2: ......
Page 6: ...CE CE ii 0086 ...
Page 14: ...viii 250 Series Maternal Fetal Monitor Revision D 2020551 001 ...
Page 16: ...1 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 29: ...Revision D 250 Series Maternal Fetal Monitor 2 1 2020551 001 2 Equipment Overview ...
Page 30: ...2 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 55: ...3 1 250 Series Maternal Fetal Monitor Revision D 2020551 001 3 Installation ...
Page 56: ...3 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 81: ...Revision D 250 Series Maternal Fetal Monitor 4 1 2020551 001 4 Maintenance ...
Page 82: ...4 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 143: ...Revision D 250 Series Maternal Fetal Monitor 5 1 2020551 001 5 Troubleshooting ...
Page 144: ...5 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 196: ...6 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 239: ...Revision D 250 Series Maternal Fetal Monitor B 1 2020551 001 B Alarms Summary ...
Page 240: ...B 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Alarms Summary For your notes ...
Page 243: ...Revision D 250 Series Maternal Fetal Monitor C 1 2020551 001 C Electromagnetic Compatibility ...
Page 251: ......