4-56
250 Series Maternal/Fetal Monitor
Revision D
2020551-001
Maintenance:
Recorder Photosensor Calibration
Adjusting the Paper-Loading Sensor
1. Load paper in the recorder. Ensure that there are
no
black squares showing to
indicate a paper-low condition. You must load at least nine sheets of paper.
2. Using a digital voltmeter, connect the positive lead to J9 (pin 6) on the Recorder
Board; connect the negative lead to J9 (pin 2).
3. Adjust R41 for a reading of +190 mVdc ± 5 mVdc on the digital voltmeter.
4. Although a factory-only fixture is required for this step, you may wish to try it
by estimating the required distance. Load paper
backwards
with the white
surface 0.380 in away from the paper tray. This creates a paper-loading error
condition. Verify that the voltage at J9 (pin 6) on the Recorder Board is at least
2.70 Vdc.
Summary of Contents for Corometrics 250 Series
Page 2: ......
Page 6: ...CE CE ii 0086 ...
Page 14: ...viii 250 Series Maternal Fetal Monitor Revision D 2020551 001 ...
Page 16: ...1 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 29: ...Revision D 250 Series Maternal Fetal Monitor 2 1 2020551 001 2 Equipment Overview ...
Page 30: ...2 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 55: ...3 1 250 Series Maternal Fetal Monitor Revision D 2020551 001 3 Installation ...
Page 56: ...3 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 81: ...Revision D 250 Series Maternal Fetal Monitor 4 1 2020551 001 4 Maintenance ...
Page 82: ...4 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 143: ...Revision D 250 Series Maternal Fetal Monitor 5 1 2020551 001 5 Troubleshooting ...
Page 144: ...5 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 196: ...6 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 239: ...Revision D 250 Series Maternal Fetal Monitor B 1 2020551 001 B Alarms Summary ...
Page 240: ...B 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Alarms Summary For your notes ...
Page 243: ...Revision D 250 Series Maternal Fetal Monitor C 1 2020551 001 C Electromagnetic Compatibility ...
Page 251: ......