4-12
250 Series Maternal/Fetal Monitor
Revision D
2020551-001
Maintenance:
Electrical Safety Tests
Patient-to-Line Leakage for ECG
1. Connect an ECG test body to the monitor’s front panel
ECG
input. The ECG test
body can be the assembly leakage cable.
2. Set the fixture’s
Leakage
switch to
Line
.
3. Turn ON the monitor using the monitor’s power switch.
4. Verify the following:
< 50
µ
A (100, 120, 220, 230, or 240 VAC monitors)
5. Record the patient-to-ground leakage for the following conditions:
Patient-to-Ground Leakage for IUP
1. Connect an IUP test body to the monitor’s front panel
UA
input. The IUP test
body can be a SensorTip transducer (wrapped in aluminum foil) or the assembly
leakage cable.
2. Set the fixture’s
Leakage
switch to
Patient
.
3. Turn ON the monitor using the monitor’s power switch.
4. Verify the following:
< 50
µ
A (100, 120, 220, 230, or 240 VAC monitors)
5. Record the patient-to-ground leakage for the following conditions:
Table 3. Patient-to-Line Leakage Conditions for ECG
Neutral
Ground
Power (Polarity)
Closed
Closed
Normal
Closed
Closed
Reversed
Open
Closed
Normal
Open
Closed
Reversed
Table 4. Patient-to-Ground Leakage Conditions for IUP
Neutral
Ground
Power (Polarity)
Closed
Closed
Normal
Closed
Open
Normal
Closed
Open
Reversed
Closed
Closed
Reversed
Open
Closed
Normal
Open
Closed
Reversed
Summary of Contents for Corometrics 250 Series
Page 2: ......
Page 6: ...CE CE ii 0086 ...
Page 14: ...viii 250 Series Maternal Fetal Monitor Revision D 2020551 001 ...
Page 16: ...1 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 29: ...Revision D 250 Series Maternal Fetal Monitor 2 1 2020551 001 2 Equipment Overview ...
Page 30: ...2 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 55: ...3 1 250 Series Maternal Fetal Monitor Revision D 2020551 001 3 Installation ...
Page 56: ...3 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 81: ...Revision D 250 Series Maternal Fetal Monitor 4 1 2020551 001 4 Maintenance ...
Page 82: ...4 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 143: ...Revision D 250 Series Maternal Fetal Monitor 5 1 2020551 001 5 Troubleshooting ...
Page 144: ...5 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 196: ...6 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 239: ...Revision D 250 Series Maternal Fetal Monitor B 1 2020551 001 B Alarms Summary ...
Page 240: ...B 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Alarms Summary For your notes ...
Page 243: ...Revision D 250 Series Maternal Fetal Monitor C 1 2020551 001 C Electromagnetic Compatibility ...
Page 251: ......