Weight
Unpacked, without input and output
table
45,0 kg
Unpacked, with input and output
table 43 cm
49,4 kg
Unpacked, with input and output
table 43 cm, incl. light cover
50,5 kg
Pixel size
High resolution
35 µm
Medium resolution
75 µm
Electrical Connection
Operating voltage
Auto ranging power supply from: 90 V to 264 V,
ac 10%
Class I with protection earth
Connect to earthed supply circuitry only.
Mains frequency
50/60 Hz
Current rating
Max. 2 A
Mains fuse protection
Europe: min. 10 A, max. 16 A
USA & Japan: min. 10 A, max. 15 A
Digitizer input voltage / current
24 V DC / max. 6,25 A
Network connectivity
Ethernet connector
RJ45 female, 10/100 Mbit/s autosensing,
shielding CAT5
Power consumption
Standby 110 V - 240 V / 50-60 Hz
configuration
max. 22 W
During operation 110 V - 240 V / 50-
60 Hz configuration
max. 140 W
Environmental conditions CRxVision during operation
Room temperature
recommended: 20 °C - 25 °C
allowed: 15 °C - 30 °C
Maximum temperature change
< 0.2 °C/min.
Relative humidity
recommended: 30 % - 60 %
allowed: 10 % - 85 %, non condensed
Light conditions
Must be operated in a dark room.
CRxVision Operating Manual V1.2
43
Summary of Contents for CRxVision
Page 2: ...Table of Contents 2 CRxVision Operating Manual V1 2 ...
Page 5: ...1 Revision History CRxVision Operating Manual V1 2 5 ...
Page 7: ...2 Introduction to this Manual CRxVision Operating Manual V1 2 7 ...
Page 26: ...4 Getting started with the CRxVision 26 CRxVision Operating Manual V1 2 ...
Page 32: ...5 Operating the CRxVision 32 CRxVision Operating Manual V1 2 ...