Aisys
4-24
04/08 M1046983
4.5.6 Bleed Resistor
The Bleed Resistor instructions appear when the focus is on the Bleed
Resistor menu item.
Instructions
The calibration fails if the flow required to reach 91 cmH2O is > 16 l/min.
• If the outcome of the calibration is Pass, the new calibration data is saved.
• If the outcome is Fail, the old calibration data is retained.
• The result of the calibration is saved to the Event Log.
Selecting Previous Menu before the calibration is done aborts the calibration
in progress and keep the old calibration constants.
Bleed Resistor Menu
Bleed Resistor Data menu
The Bleed Resistor Data menu contains a table of 17 entries from a previous
calibration. The table is erased at the start of the calibration. The table is
updated in real time during the calibration.
Read all steps before you start:
1. Complete the Insp Flow Valve calibration.
2. Put #2 plugs in the manifold and the drive gas ports of the vent engine.
3. Push the ComWheel to show the next menu.
4. Select Start.
Menu Item
Message text
Comments
Start
Calibrate manifold pressure to
bleed resistor flow.
Blocked text:
“Connect a supply of the drive
gas to continue.”
Blocked if the
ventilator drive gas
supply pressure would
cause a gas supply
failure alarm during
normal operation.
Bleed Resistor Data
Show bleed resistor calibration
table.
Previous Menu
Return to the previous menu.
During Calibration:
“Calibration in progress. Push
ComWheel to cancel.”
Summary of Contents for Datex-Ohmeda Aisys Carestation
Page 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Page 16: ...Notes xiv 04 08 M1046983 ...
Page 32: ...Notes 1 16 04 08 M1046983 ...
Page 158: ...Notes 5 20 04 08 M1046983 ...
Page 288: ...Notes 7 106 04 08 M1046983 ...
Page 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Page 356: ...Notes 9 60 04 08 M1046983 ...
Page 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Page 452: ...Notes 10 96 04 08 M1046983 ...
Page 466: ...11 14 04 08 M1046983 Notes ...
Page 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Page 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Page 497: ......