4 Install/Service Menus
M1046983 04/08
4-27
4.5.9 Cal Config
Before calibration, you must verify that the Ventilator Drive Gas and the
Altitude settings are set appropriately to match the current drive gas
configuration and machine location.
If you change any of the settings in the Cal Config menu, you must restart the
system.
Cal Config menu
Menu Item
Message text
Values
Ventilator Drive Gas
Change drive gas to match
machine configuration.
Air, O2
Altitude
Change altitude used for gas
calculations.
–400 to 3000 m
(in 100-m increments)