7 Troubleshooting
M1046983 04/08
7-75
VENT +12.5V FAIL
Vent Fail. Monitoring
Only
Nominal 12.5V <11.3 Vdc or
>13.13 Vdc
Medium
AC,
Vent
In the Service Software / Vent Interface Bd Power Diagnosis, view the “Vent Int Bd 10VA Voltage” from
Board Supplies:
If “Vent Int Bd 10VA Voltage” reads “OK, and +12.5 Vdc reads “Fail”, replace the VIB.
If “Vent Int Bd 10VA Voltage” reads “Fail” and the “Vent Int Bd 10VA Voltage” from the Anes Cntrl Bd
reads “OK”, Check cabling between ACB and VIB.
VENT +6V FAIL
Vent Fail. Monitoring
Only
VSIB +6V out of range<5.51
Vdc or > 6.5 Vdc
High
AC,
Vent
Vent +12.5 V (10 VA) is
OK
In the Service Software / Vent Interface Bd Power Diagnosis, view the “Vent Int Bd 10VA Voltage” from
Board Supplies:
If “Vent Int Bd 10VA Voltage” reads “OK, and +6.0Vdc reads “Fail”, replace the VIB.
If “Vent Int Bd 10VA Voltage” reads “Fail” and the “Vent Int Bd 10VA Voltage” from the Anes Cntrl Bd
reads “OK”, Check cabling between ACB and VIB.
VENT 1.22V FAIL
Vent Fail. Monitoring
Only
Voltage < 1.074Vdc or Voltage
> 1.367 Vdc
Medium
AC,
Vent
Vent +12.5 V (10 VA) is
OK
In the Service Software / Vent Interface Bd Power Diagnosis, view the “Vent Int Bd 10VA Voltage” from
Board Supplies:
If “Vent Int Bd 10VA Voltage” reads “OK, and 1.22 Vdc reads “Fail”, replace the VIB.
If “Vent Int Bd 10VA Voltage” reads “Fail” and the “Vent Int Bd 10VA Voltage” from the Anes Cntrl Bd
reads “OK”, Check cabling between ACB and VIB.
VENT -6V FAIL
Vent Fail. Monitoring
Only
VSIB -6V out of range<-6.72
Vdc or > -5.28 Vdc
High
AC,
Vent
Vent +12.5 V (10 VA) is
OK
In the Service Software / Vent Interface Bd Power Diagnosis, view the “Vent Int Bd 10VA Voltage” from
Board Supplies:
If “Vent Int Bd 10VA Voltage” reads “OK, and -6.0Vdc reads “Fail”, replace the VIB.
If “Vent Int Bd 10VA Voltage” reads “Fail” and the “Vent Int Bd 10VA Voltage” from the Anes Cntrl Bd
reads “OK”, Check cabling between ACB and VIB.
VENT ADC VREF FAIL
Vent Fail. Monitoring
Only
VSIB ADC3.200V ref voltage
out of range <3.179 or
>3.221 Vdc
High
AC,
Vent
Vent +12.5 V (10 VA) is
OK
In the Service Software / Vent Interface Bd Power Diagnosis, view the “Vent Int Bd 10VA Voltage” from
Board Supplies:
If “Vent Int Bd 10VA Voltage” reads “OK, and 3.2 Vdc reads “Fail”, replace the VIB.
If “Vent Int Bd 10VA Voltage” reads “Fail” and the “Vent Int Bd 10VA Voltage” from the Anes Cntrl Bd
reads “OK”, Check cabling between ACB and VIB.
VENT AIRWAY
OVERPRESS SIGNAL
Inspiration stopped
High airway overpressure
signal set.
Medium
AC,
Vent
Mechanical Ventilation
On
No Service Action.
Reboot system. If problem continues, check Airway Pressure signal in Service Mode.
VENT AIRWAY
OVERPRESS SIGNAL
FAIL
Vent Fail. Monitoring
Only
Ventilator SIB indicates the
High Airway overpressure
signal was set and Paw < 90
cmH
2
O and Pmanifold <80
cm H
2
O.
Medium
AC,
Vent
Mechanical Ventilation
On
No Service Action.
Reboot system. If problem continues, check Airway Pressure signal in Service Mode.
Error Log Entry
Alarm Text
Condition (Basic info)
Priority Source Enabling Criteria
Action/Troubleshooting
Summary of Contents for Datex-Ohmeda Aisys Carestation
Page 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Page 16: ...Notes xiv 04 08 M1046983 ...
Page 32: ...Notes 1 16 04 08 M1046983 ...
Page 158: ...Notes 5 20 04 08 M1046983 ...
Page 288: ...Notes 7 106 04 08 M1046983 ...
Page 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Page 356: ...Notes 9 60 04 08 M1046983 ...
Page 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Page 452: ...Notes 10 96 04 08 M1046983 ...
Page 466: ...11 14 04 08 M1046983 Notes ...
Page 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Page 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Page 497: ......