Aisys
7-88
04/08 M1046983
7.10.8 eVap Communication troubleshooting
Flowchart
23
Problem Continues?
Does either LED indicate
Activity during Start-up?
Does the eVap communicate
in the software download mode?
Start
Yes
Perform Software
Download.
Check the ADB
TXD / RXD LEDs
Can the issue be
duplicated?
TXD
RXD Only or No Activity
No
Contact Technical
Support.
No
Check / Replace
ACB to Pan Cable
and Pan to eVap
Cable.
Problem Solved
Yes
Check / Replace
Cabling. Replace
ACB if problem
persists.
Replace ADB
Board.
Is the Mixer functioning?
No
Troubleshoot
Mixer.
Yes
No
Problem Continues?
Replace the PAN
Connector Board.
Yes
No
Error Log Entry
COM ERROR VENT TO ACB
Compatibility failure: DCB hardware
VAPORIZER LOST MIXER FLOW
VAPORIZER SUBSYSTEM COMM FAILURE
Yes
Symptom
AB.75.180
Summary of Contents for Datex-Ohmeda Aisys Carestation
Page 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Page 16: ...Notes xiv 04 08 M1046983 ...
Page 32: ...Notes 1 16 04 08 M1046983 ...
Page 158: ...Notes 5 20 04 08 M1046983 ...
Page 288: ...Notes 7 106 04 08 M1046983 ...
Page 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Page 356: ...Notes 9 60 04 08 M1046983 ...
Page 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Page 452: ...Notes 10 96 04 08 M1046983 ...
Page 466: ...11 14 04 08 M1046983 Notes ...
Page 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Page 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Page 497: ......