Aisys
9-12
04/08 M1046983
9.7.1 Remove the
electronic
vaporizer
1. If the eVap is equipped with bulkhead fittings on the eVap support bracket,
disconnect the inlet, outlet, and scavenging tubing from the bulkhead fittings.
Remove the suction tubing connectors (if equipped) from the clips on the bottom of
the eVap support bracket.
If the eVap is not equipped with bulkhead fittings, disconnect the tubing directly from
the inlet, outlet, and the scavenging ports of the eVap.
2. Disconnect the pan connector harness (
A
) from the Agent Delivery board.
3. Clip the tie-wrap holding the pan connector harness to the eVap support bracket.
4. Remove the six screws (
B
) holding the eVap to the dashboard.
5.
Note
: The eVap assembly weighs approximately 5 kg (11 lb). Most of the weight of
the eVap is at the rear.
w
CAUTION
When handling the eVap, be careful not to damage the Flow Control Valve at the
back of the flowmeter block.
6. To remove the eVap:
• Firmly take hold of the eVap at the rear with one hand without touching any
boards.
• Support the front of the eVap with the other hand.
• Raise the rear of the eVap tilting the dashboard forward.
• When clear, remove the eVap from the machine.
Scavenging
Outlet
Inlet
Suction Tubing Clips
(A)
(B)
(B)
Summary of Contents for Datex-Ohmeda Aisys Carestation
Page 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Page 16: ...Notes xiv 04 08 M1046983 ...
Page 32: ...Notes 1 16 04 08 M1046983 ...
Page 158: ...Notes 5 20 04 08 M1046983 ...
Page 288: ...Notes 7 106 04 08 M1046983 ...
Page 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Page 356: ...Notes 9 60 04 08 M1046983 ...
Page 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Page 452: ...Notes 10 96 04 08 M1046983 ...
Page 466: ...11 14 04 08 M1046983 Notes ...
Page 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Page 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Page 497: ......