2.8 Ventilator Interface board
The Ventilator Interface board (
A
) provides the electrical and/or pneumatic interface to
the following:
• Inspiratory (
B
) and expiratory (
C
) flow sensors (transducers)
• Patient airway (
D
) and manifold (
E
) pressures (transducers)
• Oxygen sensor (in breathing system)
• Inspiratory Flow Valve
• Gas Inlet Valve
• SCGO solenoid, SCGO/CGO position switches (if SCGO installed)
• Canister release switch
• ABS On switch
• ACGO position switch (if ACGO installed)
• O
2
Flush switch
• Bag/Vent switch
• Accessory Power (for task lights)
The Ventilator Interface board functions are managed locally by a microcontroller. The
microcontroller communicates data values to the controlling CPU via an RS-422 serial
interface.
Figure 2-17 • Ventilator Interface board
A
B
C
E
D
Summary of Contents for Datex-Ohmeda Aisys Carestation
Page 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Page 16: ...Notes xiv 04 08 M1046983 ...
Page 32: ...Notes 1 16 04 08 M1046983 ...
Page 158: ...Notes 5 20 04 08 M1046983 ...
Page 288: ...Notes 7 106 04 08 M1046983 ...
Page 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Page 356: ...Notes 9 60 04 08 M1046983 ...
Page 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Page 452: ...Notes 10 96 04 08 M1046983 ...
Page 466: ...11 14 04 08 M1046983 Notes ...
Page 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Page 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Page 497: ......