2 Theory of Operation
M1046983 04/08
2-47
Manual expiration
The Bag/Vent switch keeps the ventilator path closed (
B
).
Gas flows from the patient (
4
), through a unidirectional valve (expiratory check valve),
and into the bag (
5
).
During exhalation, fresh gas flows backwards through the absorber (
FG
) into the
expiratory limb, downstream of the expiratory check valve.
For machines that are plumbed to return sample gas to the breathing system, the
returned gas (
SGR
) enters the breathing system after the expiratory check valve.
Figure 2-34 • Flow during manual expiration
AP
Airway Pressure
B
Bag/Vent switch to Bag
FG
Fresh Gas
SGR
Sample Gas Return
4
Expiratory flow
5
Flow to bag
FG
AP
5
FG
AB.82.027
SGR
4
5
B