Aisys
3-2
04/08 M1046983
3.1 Inspect the system
w
w
w
w
CAUTION
The upper shelf weight limit is 45 kg (100 lb).
w
w
w
w
WARNING
Do not leave gas cylinder valves open if the pipeline supply is in use.
Cylinder supplies could be depleted, leaving an insufficient reserve
supply in case of pipeline failure
.
Before testing the system, ensure that:
•
The equipment is not damaged.
•
Components are correctly attached.
•
The breathing circuit is correctly connected, not damaged.
•
Pipeline gas supplies are connected.
•
Cylinder valves are closed.
•
Models with cylinder supplies have a cylinder wrench attached to the
system.
•
Models with cylinder supplies have a reserve supply of O
2
connected to
the machine during system checkout.
•
The casters are not loose and the brakes are set and prevent movement.
•
The power cord is connected to a wall outlet. The mains indicator comes
on when AC Power is connected.
3.2 System checkout
3.2.1 Leak < 250 ml
The
Leak < 250 ml
setting is used during the circuit leak check portion of the
checkout procedures. This check tests for leaks in the machine, breathing
circuit, patient circuit, and manual bag. The default setting is
No
.
Note
Extraction of gas by external gas monitors may cause failure of the leak checks
during tests.
When No is selected, the leak test will pass for leaks below 250 ml at 3 kPa
(30 cmH
2
O) pressure with no user interaction required. For leaks between
250 ml and 750 ml, the user can fix the leak and rerun the test or accept the
leak and continue. For leaks above 750 ml, the test will fail and the user must
fix the leak and rerun the test.
Set to
Yes
to quantify small leaks above 100 ml during the checkout
procedures. Selecting
Yes
will display the measured leak at 3 kPa (30
cmH
2
O) pressure and result in the test taking somewhat longer.
Summary of Contents for Datex-Ohmeda Aisys Carestation
Page 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Page 16: ...Notes xiv 04 08 M1046983 ...
Page 32: ...Notes 1 16 04 08 M1046983 ...
Page 158: ...Notes 5 20 04 08 M1046983 ...
Page 288: ...Notes 7 106 04 08 M1046983 ...
Page 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Page 356: ...Notes 9 60 04 08 M1046983 ...
Page 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Page 452: ...Notes 10 96 04 08 M1046983 ...
Page 466: ...11 14 04 08 M1046983 Notes ...
Page 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Page 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Page 497: ......