3.12 Power failure test
1. Connect the power cord to a wall outlet. The mains indicator on the front
panel comes on when AC Power is connected.
2. Set the system switch to On and Start a case.
3. Unplug the power cord with the system turned on.
4. Make sure that the power failure alarm comes on.
5. Make sure the following message is displayed:
• Plug in power cable. On battery
6. Connect the power cable again.
7. Make sure the alarm cancels.
3.13 Electrical safety tests
Make sure the system is completely assembled and all accessory devices are
connected to electrical outlets.
1. Connect an approved test device (e.g. UL, CSA, or AAMI) and verify that the
leakage current is less than:
2. Make sure that the resistance to ground is less than 0.2
Ω
between an
exposed metal surface and the ground pin on the power cord.
AB.75.052
Voltage
Max. Leakage Current
120/100 Vac
300
µ
Amps
220/240 Vac
500
µ
Amps
Summary of Contents for Datex-Ohmeda Aisys Carestation
Page 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Page 16: ...Notes xiv 04 08 M1046983 ...
Page 32: ...Notes 1 16 04 08 M1046983 ...
Page 158: ...Notes 5 20 04 08 M1046983 ...
Page 288: ...Notes 7 106 04 08 M1046983 ...
Page 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Page 356: ...Notes 9 60 04 08 M1046983 ...
Page 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Page 452: ...Notes 10 96 04 08 M1046983 ...
Page 466: ...11 14 04 08 M1046983 Notes ...
Page 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Page 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Page 497: ......