6.12.6 Right-side cable route
To access the cable route on the right side of the machine,
remove the lower and upper right-side cosmetic panels.
Route the cables through the cable clamps shown
(
arrows
).
Note:
When replacing the lower cosmetic panel, position
the cables within the beveled edge of the extrusion to
allow a flush mount of the panel without pinching the
cables
Cable from the upper chassis can be routed in this area
through the cutout (
T
) in the chassis.
Additional cables can be routed into this area through the
upper cosmetic panel. First, remove the plug (
U
); then,
route the cable through the upper panel. Flip the plug over
before replacing it onto the cosmetic panel.
T
U