Aisys
7-34
04/08 M1046983
Alternate O
2
/ System Malfunction Screen Troubleshooting
Flowchart
11
AB.91.013
Did Customer
depress the
Button?
Are any of the entries
contained in the
analyzed log?
Start
Compare the
Combined log to
the Error Logs in
the Table in this
document.
No
Yes
Follow the appropriate
Flowchart for the error listed.
Check the Alt O2 Button and
wiring harness, ABS Filter
Connections, and ACB
Connections.
Is “ACComm: Entering Alt-O2 Gas
Mode” in the analyzed log around
the Date and Time of the issue?
Verify
with Customer
No
Yes
No
Normal Operation.
No Service Issue.
Yes
Analyze
the
combined
log.
Download and combine the
Error / Alarm Logs.
AB.75.075